![Amsino AMSure User Manual Download Page 62](http://html1.mh-extra.com/html/amsino/amsure/amsure_user-manual_2936262062.webp)
File
Name
IFU of
AMSure
®
Enteral
Feeding
Pump and Feeding Set
REF
#
L10196_US
Version
B
Page
62
of
81
AMSINO Confidential
8.11 Continuous beeping
sound lasting
approximately 3 mins.
Total Power Failure
Stops the feeding
Pump turns off.
Potential Cause:
•
Battery and power adapter
failure.
•
Connect the pump
to AC power supply
immediately and
restart the pump.
8.12
Motor Error
Stops the feeding
Potential Cause:
•
The motor is malfunction.
•
Please contact
customer service.
8.13
Communication Fault
Stops the feeding
Potential Cause:
•
System error for processor.
•
Please contact