left eyepiece clear as same as the right eyepiece. 

4.5 Focusing adjustment 

Usually 10X objective should be used for focusing adjustment first. Take 10X 

objective into bright path, then observe the specimen with right eyepiece by right eye 

and turn coarse focusing knob to find the image, then turn slowly fine focusing knob 

to make the image clear. 

All objectives of our microscope are precisely adjusted, and approximate clear image 

should be found with the other objectives after getting the clear image with 10X 

objective. If the image isn’t enough clear with the other objectives, please turn fine 

focusing knob slightly to find the satisfactory image. 

Don’t turn left and right coarse and fine focusing knobs with different direction with 

power at the same time, if so, the focusing system will be damaged. 

Don’t directly pull objective to turn nosepiece when user changes the different 

objectives, if so,    optical quality of 

microscope possibly be affected. The 

right way is to take the tooth-like part 

of the nosepiece to turn it, and make 

the objective into correct position and 

into the bright path.

 

 

4.6 Usage of the tension adjustment 

ring (Shown in Fig.6) 

The tension adjustment ring can adjust the tension of the coarse and fine focusing 

unit to prevent the stage from sliding down automatically and to improve the comfort 

Fig.6  Coarse  A   F

F

 

nd  ine  ocusing

Left Coarse 

Left Fine 

Right Fine 

Right Coarse 

Tension 

Focusing Knob

Focusing Knob

Focusing Knob

 Adjustment Ring

Focusing Knob

 Knobs And Tension Adjustment Ring

 

Summary of Contents for B100

Page 1: ...15 B100 B200 Series Biological Microscope Operation Instruction To ensure the safety and obtain satisfactory performance please study this operation instruction thoroughly before your operation ...

Page 2: ...16 Contents 1 Purpose 2 2 Specifications 2 3 Installation 4 4 Operation 4 5 Maintenance 8 6 Troubleshooting 10 7 Outfits 13 ...

Page 3: ... Handleof IrisAperture Pow er S witchAnd Light Collector Fig 1 B200Binocular Biological Microscope FineFocusingKnob Light Controller Condenser Up downKnob Coaxial CoarseAnd ElectricBase Eyepiece Nosepiece O bjective FineFocusingKnob Monocular Head MainBody Clamp S tage Condenser Light Controller Power S witchAnd Light Controller Fig 2 B100Monocular Biological Microscope ...

Page 4: ...d family 2 Specifications 2 1 Total magnifications Objective Total Magnifications Eyepiece 4X 10X 40X 100X 10X 40X 100X 400X 1000X 16X 64X 160X 640X 1600X 2 2 Objectives Objective Numerical Aperture N A Objective Field Resolving Power Working Distance Field Number ф18 Field Number ф20 Achromatic Objective Plan Objective 4X 0 10 4 5mm 5mm 2 8µm 13mm 16mm 10X 0 25 1 8mm 2mm 1 1µm 6 3mm 2mm 40X S 0 6...

Page 5: ... Mechanical stage Size 142mm 134mm X Y travel 50mm 70mm 2 3 6 Focusing systems Coaxial coarse and fine focusing knobs Coarse stroke 22mm Fine division 2µm Condenser up down range 25mm 2 3 7 Condenser Abbe condenser N A 1 25 Adjustable aperture Aperture center is adjustable 2 3 8 Mirror Mirror diameterф52mm optional 2 3 9 Filter Built in blue filer ф38 5mm 2 3 10 Anti fungal systems Eyepiece and ob...

Page 6: ...king box and drive them into the holes of nosepiece orderly and tightly according to the times Take out the oil bottle and dust cover and place it properly for convenient use 4 Operation 4 1 Electric components Make sure the supply voltage meets the instrument s requirement then plug in it Rotation potentiometer with power switch used by the instrument is for protecting lamp Mirror Bracket Mirror ...

Page 7: ...st size of iris aperture for adjusting the contrast of the specimen 4 4 Interpupillary distance and diopter adjustment Shown in Fig 5 4 4 1 Interpupillar distance is different for everyone so interpupillary distance should be adjusted before using binocular microscope Please take the two eyepiece tubes to turn until the bright rings observed by two eyes are in superposition entirely B100 monocular...

Page 8: ...n t turn left and right coarse and fine focusing knobs with different direction with power at the same time if so the focusing system will be damaged Don t directly pull objective to turn nosepiece when user changes the different objectives if so optical quality of microscope possibly be affected The right way is to take the tooth like part of the nosepiece to turn it and make the objective into c...

Page 9: ...nto bright path At this time nosepiece or stage moving knob should be turned slightly and also slightly turn fine focusing knob to clear away the air bubble of immersion oil otherwise the air bubble would affect the image badly After immersion oil used the oil of specimen and the microscope surface should be soon cleaned by absorbent cotton lens paper gauze or soft cotton cloth with moderate mixtu...

Page 10: ...on Finally cover microscope with dust cover 4 9 2 Please take eyepiece and objectives out from microscope if user will stop using it for a long time and place eyepiece and objectives into drier with drying agent Cover microscope with dust cover 5 Maintenance 5 1 Clean microscope 5 1 1 Don t touch the lens with hand Dust on lens should be cleaned by soft brush or absorbent cotton or cleaned by abso...

Page 11: ...s and mist damage instrument 5 2 4 Please pay attention to prevent microscope from violent shake and vibration in application and in carrying Don t drag it on the surface of worktable to avoid damage microscope and worktable U screwKnurled n Lam pAdjustm ent Knurled Pull O ver Bulb HousingC over ReplaceBulb C loseC over TheLam p C over Thum bS crew Thum bS crew Fig 7 Replacement of Bulb 5 3 Replac...

Page 12: ...ly better cut off the microscope about 30 minutes 5 4 Replacement of fuse Shown in Fig 8 5 4 1 Cut off power of microscope and pull out the plug 5 4 2 Unscrew fuse cap in the back of base take out old fuse 5 4 3 Replace a new fuse then screw the fuse cap 5 5 Stop to use microscope please cut off power cover the dust cover and place it in a cool and dry environment 6 Troubleshooting In the period o...

Page 13: ...uirement Replacing bulb Brightness isn t adjusted correctly Adjust rotation potentiometer Objective isn t in correct position Make the objective in correct position The size of iris aperture is too small Adjust the size of iris aperture Lens objective eyepiece condenser light collector has dust Clean it Position of condenser is too low Higher condenser Image isn t clear Cover glass of specimen doe...

Page 14: ... of image is dark or image is moving as focusing Objective isn t in correct position Make the objective in correct position Specimen isn t placed correctly Place specimen levelly on stage and clip it with clamp Objective touches specimen as changing low times objective to high times objective Cover glass of specimen isn t in up direction Place specimen correctly Cover glass doesn t meet the requir...

Page 15: ...10 B A4 10X 0 25 B A10 40X 0 65 S B A40 100X 1 25 Oil S B AO100 Plan Objective 4X 0 10 B P4 10X 0 25 B P10 40X 0 65 S B P40 100X 1 25 Oil S B PO100 Monocular Head Inclined 30 Rotatable 360 Mo 1 Seidentopf Binocular Head Inclined 30 Rotatable 360 Interpupilary Distance 55 75mm Bi 1 Nosepiece Quadplex Coaxial Coarse and Fine Focusing Knobs N 4 Fixed Stage Clips S 1 Attachable Mechanical Stage Travel...

Page 16: ...lt In Blue Filter Mo Incandescent Lamp 110V 220V 20W Spare Lamp 1pc Built In Blue Filter E1 Halogen Lamp 6V 20W Wide Voltage 85V 265V Spare Lamp 1 pc Built In Blue Filter E2 LED Illumination Systems LED Fuse Spare Fuse 2 pcs Note In Table Is Standard Attachment Is Optional Accessories Packing Size 340mm X 243mm X 425mm Gross Weight 7 kgs Net Weight 5 5 kgs ...

Reviews: