![Amsco 36 H Operator'S Manual Download Page 134](http://html1.mh-extra.com/html/amsco/36-h/36-h_operators-manual_2936188134.webp)
9-18
P129394-068
Operator Manual
Troubleshooting
Alarm
Description
Screen with Operator Instructions
9.4.4 Too Long To
Unseal Door
Occurs if door seal
pressure does not drop
below 5 psig within
allotted time.
Screen with
Service
Instructions
Screen with Operator Instructions
9.4.5 Chamber
Pressure Transducer
Failure
Occurs if chamber
pressure reading is
outside the normal
range.
STATUS . . ALARM!
229
TOO LONG TO UNSEAL DOOR
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
DOOR IS UNSEALED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
3. IF LOAD MUST BE REMOVED, REFER TO
EMERGENCY DOOR OPERATION PROCEDURE
IN OPERATING MANUAL
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
230
TOO LONG TO UNSEAL DOOR
-> DOOR SEAL PRESSURE NOT BELOW 10 PSIG
WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. SEAL PRESSURE SWITCH MALFUNCTION
-> READJUST PS1 (PS2)
-> REPAIR PS1 (PS2)
2. SEAL NOT RETRACTING
-> CHECK SEAL STEAM
-> CHECK SEAL EXHAUST
3. SOLENOID VALVE MALFUNCTION
-> REPAIR S37 (S38)
-> REPAIR S35 (S36)
PAPER
FEED
EXIT
STATUS . . ALARM!
217
CHAMBER PRESSURE TRANSDUCER FAIL-
URE
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY ABORT CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
Summary of Contents for 36 H
Page 2: ......
Page 14: ...xii P129394 068 Operator Manual Techniques of Sterilization LIST OF FIGURES Descrption Page ...
Page 59: ...6 11 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 63: ...6 15 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 67: ...6 19 Sterilizer Operation Operator Manual P129394 068 Healthcare Capital Equipment ...
Page 78: ...6 30 P129394 068 Operator Manual Sterilizer Operation Healthcare Capital Equipment ...
Page 116: ...8 10 P129394 068 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 140: ...9 24 P129394 068 Operator Manual Troubleshooting Healthcare Capital Equipment ...
Page 152: ...10 12 P129394 068 Operator Manual Service Procedures Healthcare Capital Equipment ...