Section 7 Preventive Maintenance
Copyright 2009, AMS, Proprietary Information, All Rights Reserved
7-3
P/N 0010-0667 , Rev A
Below are the standard measurement tools for calibration.
AMS
part
#
Model
Manufacturer Calibration Required
5400-0298
Power Meter Nova 2
5400-0300
Detector 633
7Z02410 / PD300-SH
5400-0299
Detector 523 / 250
Watt7Z02676S
FL250A-LP1-SH-V1
5400-0302
L30A-10MM
7Z02273
Ophir Optronic Inc. 2
Electronics Ave., Danvers
Industrial Park Danvers Ma
01923 Phone 435-753-3729
Yearly
5400-0294
43 Thruline®
Wattmeter
www.bird-electronic.com
5400-0296
Watt Element -
100A/25-60 MHz/
100 Watt Elements
or
5400-0295 each
Quick Change
Solderless" RF
Connectors #4240-
125
Yearly
0133-4471/50
Hz
Flow Meter Model S-
111 European
AMS
Yearly
0133-4470/60
Hz
Flow Meter Model S-
111 Domestic 110V
AMS
Yearly
Power Displayed
Acceptable Range
20 Watts
19 - 23 watts
40 Watts
38 - 45 Watts
60 Watts
58 - 68 Watts
80 Watts
78 - 91 watts
100 Watts
98 - 113 Watts
120 Watts
119 - 136 Watts
140 Watts
139
-
158 Watts
160 Watts
159 - 181 Watts
180 Watts
179 - 198 Watts
Summary of Contents for GreenLight XPS