15
Influence of
external low
frequency
magnetic fields
(E11)
10 A/m: add ±0.1 mA
30 A/m: add ±0.3 mA
@ I
N
: 3.500 mA – 600.0 mA / 40 Hz
to 1 kHz
and a frequency of the magnetic
field of 15 to 400 Hz acc. to
IEC 61000-4-8
Influence of load
current (E12)
Add ±6 μA per A load current
Influence of touch
current caused
common mode
voltage (E13)
N/A
Influence of
frequency (E14)
N/A
Influence of
repeatability (E15)
N/A
Agency approval
Power supply
Two 1.5 V AAA (LR03) batteries
Battery life
60 hours typical
Low battery
voltage
Approx. 2.5 V
Low battery
indication
Auto power off
Idle for 20 minutes
Dimensions
(H x W x L)
8.7 x 3.5x 1.8 in
(221 x 89 x 48 mm)
Weight
Approximately 0.90 lb (410 g) with
battery installed
ELECTRICAL SPECIFICATIONS
Accuracy is given as ± (% of r counts of least
significant digit) at 23 °C ± 2 °C (≤ 80% R.H.)
AC Function
• AC A specifications are AC coupled, true RMS.
• For non-sinusoidal waveforms, Additional Accuracy by
Crest Factor (C.F.):
Add 1.0% for C.F. 1.0 to 2.0
Add 2.5% for C.F. 2.0 to 2.5
Add 4.0% for C.F. 2.5 to 3.0
• Max. Crest Factor of Input Signal:
3.0 @ 3000 counts
2.0 @ 4500 counts
1.5 @ 6000 counts
• Frequency Response is specified for sine waveform.
• When operating under magnetic field add error E11.
• When measuring leakage current of multiple conductors,
add error E12 for influence of load current.
• For Position Error see error E1.
Summary of Contents for ALC-110
Page 1: ...ALC 110 Leakage Current Clamp User Manual ENG FRE SPA ...
Page 2: ......
Page 14: ...10 DATA HOLD Press 20A 0A MAX HOLD Press 2 sec ...
Page 22: ......
Page 34: ...10 CONSERVATION DES DONNÉES Appuyer 20A 0A CONSERVATION DU MAXIMUM Appuyer 2s ...
Page 42: ......
Page 54: ...10 MEMORIA DE DATOS Presione 20A 0A MEMORIA DE VALORES MÁXIMOS Presione 2 seg ...