USING THE OTOWAVE
6
D-0014930-A
– Otowave 102-C Operating Manual
4.
USING THE OTOWAVE
4.1.
BATTERY-PACK
The Otowave 102-C is powered by a rechargeable, built-in Nickel-Metal Hydride (NiMH) battery-pack. The
battery is not intended to be changed by the user. The battery pack may explode or cause burns, if
disassembled, crushed or exposed to fire or high temperatures. Do not short-circuit.
A battery state indicator is shown in the top right corner of the display (except when showing test results).
This shows
the battery state as a progressively emptying battery. The 102-C should always be placed on its
cradle when not in use to allow the battery to trickle charge. When
the symbol “
!”
appears next to the battery
state indicator, or when advised at switch-on the handset must be placed on its cradle to charge before further
use.
Depleated batteries do not affect the instrument configuration, the contents of the database, the calibration
settings or the results of the last test.
4.2.
OPERATING LANGUAGE
To set the operating language (English, French,Spanish, Italian, Portugese or German) use the options within
the CONFIGURATION menu.
4.3.
THE CRADLE
4.3.1.
GENERAL
The Otowave 102-C is powered by a rechargeable Nickel-Metal Hydride (NiMH) battery-pack which is fitted in
the instrument. If the instrument is placed onto its cradle the battery will trickle chage.
4.3.2.
CONNECTORS
The mains adapter is supplied as part of the equipment. Connect the output lead from the adapter into the
power socket in the rear of the instrument cradle. Switch on the mains supply. The mains adapter is the mains
disconnect device and therefore the tympanometer should be positioned such that easy access to the mains
adapter is possible.
The output from mains adapter is fitted with electronic circuit protection. In case of overload the adapter will
shut down. When the fault is cleared the adapter will operate as normal. However, the input to the mains
Power
USB (PC)
Printer