TECHNICAL SPECIFICATION
30
D-0014930-A
– Otowave 102-C Operating Manual
Backup power supply
> 30 days without the battery being recharged on its
cradle
Languages
Operating Languages
English, German, French, Spanish, Portuguese or Italian
Printing
Supported printer
Sanibel MPT-II
Interface
Wired connection to cradle
Information printed
Tympanogram, Tympanogram analysis parameters,
Reflex graphs, Reflex analysis parameters, Serial
Number of device, Last and Next Due Calibration dates;
space for patient & clinician’s details to be entered.
Interface to computer
Interface
USB version 1.1
Information sent
Patient header, left and right ear test data.
Power Supply
Battery Types
NiMH rechargeable battery pack (built-in)
Mains power (to cradle)
100-240Vac; 50/60Hz; 0.2A
Warm-up period
None at room temperature
Number of recordings without recharge
Up to 100
Auto power-off delay
90 or 180 seconds
Idle current
70mA
Current while testing
230mA
Physical
Display
128 x 64 pixels / 8 lines of 21 characters
Dimensions
230mm (L) x 115mm (W) x 70mm (H)
Total Weight (handset and cradle)
650g
Environmental
Operating temperature range
+15
o
C to +35
o
C
Operating humidity range
30% to 90% RH, non-condensing
Operating atmospheric pressure range
980 to 1040 mb
Transport and storage temperature range
-20
o
C to +70
o
C
Transport and storage humidity range
10% to 90% RH, non-condensing
Transport
and
storage
atmospheric
pressure range
900 to 1100 mb
Standards conformance
Safety
IEC 60601-1(plus UL, CSA & EN deviations)
EMC
IEC 60601-1-2
Performance
IEC 60645-5, Type 2 Tympanometer
CE mark
To the EU Medical Device Directive
15.2.
EQUIPMENT CLASSIFICATION
Type of protection against electric shock
Internally Powered
Degree of protection against electric shock
Type BF applied part
Degree of protection against ingress of water
Not protected
Mode of operation
Continuous operation
Equipment mobility
Portable
The Otowave 102-C Tympanometer is classified as a Class IIa device under Annex IX of the EU Medical
Devices Directive.