20
D-0116492-B (OM022)
– Otosure Operating Manual
To accept the results and transfer them to Audibase click on the “Accept” button. The Otosure window will
close and the Audibase window will be displayed. Refer to the Audibase user manual for additional options,
but it should be noted that the results must be saved in Audibase to be retained in the database.
To continue testing click
on the “Continue” option. The dialogue box will close and the operator may continue
testing as described in these sub-sections.
5.9.
SINGLE-FREQUENCY TESTING
In Computer test mode it is possible to run a test at a single frequency. For example, the operator might
have noticed that the patient response at a particular frequency was erratic, or the automatic test was unable
to establish a response at a particular frequency.
The “Test > One freq. test” drop-down menu option may be used to select a single frequency for a particular
ear. Clicking on the required option for frequency and ear initiates a test which will override any audiometric
data already established for that frequency. At the end of the test the option to “Continue” or “Accept” is
presented.
5.10.
MIXED-MODE TESTING
This allows an audiogram to be completed by performing tests (typically at a few selected frequencies) using
an alternative test type. For example, this facility could be used to perform a manual test if the automatic test
option was unable to establish a threshold at a particular frequency).
The legend “Mixed” is displayed at the bottom right of the Otosure window.