13
D-0116492-B (OM022)
– Otosure Operating Manual
4.
INSTALLING SOFTWARE & OPERATIONAL CHECK
4.1.
GENERAL
This section describes the procedure for installing the Otosure operating software & associated device
drivers, undertaking the first-time connection of the Otosure audiometer and performing a basic operational
check to ensure that the instrument is working.
4.2.
INSTALLATION
Installation is a two-stage process (please ensure both are completed):
1. Install the Otosure software on the PC
2. Install the device drivers to enable USB communication between the PC and the Otosure audiometer
Step 1: Install the Amplivox Otosure Software
The installation of the Otosure software is seamless and is installed as part of Audibase. Please refer to the
Audibase5.x user guide for step by step installation instructions.
Step 2: Install the USB Device Drivers
Please refer to
“D-0115682” Guide to Installing Amplivox USB Drivers
4.3.
UNINSTALLING THE OTOSURE PROGRAM
Uninstalling the Otosure program is achieved simply by navigating to the location where the software was
originally installed and deleting the file named “Otosure.exe”. There is no need to use the “Add or Remove
Programs” utility.
This will leave Audibase installed.
Please note:
If more information about the uninstallation of the Audibase are required, please refer to
the Audibase5.x user guide.
4.4.
OPERATIONAL CHECK
Connect the headphones and patient response switch to the Otosure audiometer, and connect the Otosure
audiometer to the PC via a USB cable. Start the Audibase application on the PC and enter new patient
details (refer to the Audibase user manual).
Start a test using the Launch Test button in the Audibase toolbar
Close any dialogue boxes in the Otosure window and select the Manual Test option:
Move the mouse pointer over the “Present” icon in the Otosure window and ensure a tone is heard from an
earphone. Use the arrow keys on the PC keyboard to change the level and frequency of the tone. Use the
“L” and “R” keys on the PC keyboard to switch between earphones.
Press the patient response switch and ensure that the “Response” icon in the Otosure window changes
colour to light green and returns to dark green when the switch is released.