TEST MODE
17
OM033-4 - CA850 Series 5 Operating Manual
Press the ‘This test’ softkey to delete hearing thresholds.
9.3.2.4 Cancel
Press the ‘Cancel’ softkey to cancel the erase selection and return to the audiogram page.
9.3.3 Plot point
Pressing
the ‘Plot point’ softkey will plot a hearing threshold for the specified ear, frequency and sound level.
This feature is used when performing manual audiometry in conjunction with the
and
keys (to change
frequency) and the
and
keys (to change level).
9.3.4 Next Page
Press the ‘Next Page’ softkey to access the following page within the TEST mode.
9.4
TEST DETAILS PAGE
The Test Details page is normally used to view test details numerically. However it is possible to enter data
directly on this page, for example the examiner, threshold values and the test date. Use the
and
keys to
select the required data field and enter data from the keyboard. Thresholds are always entered in the
‘Current’ columns (use ‘ALT’ ‘I’ to enter a negative value). If specific frequencies have been omitted from the
test (see Section 11.2.6) the
key may be used to skip over these data fields.
Refer to Section 9.5 for details of how to enter patient details and previous data, and then immediately
perform a new test.
SOFTKEY OPTIONS
9.4.1 Save test
Press the ‘Save test’ softkey to save the test record within the internal database. If a record with the same
patient number and date already exists, the operator will be prompted either to create a new record or overwrite
the existing one. If required, the database functions may be used to confirm the save operation (see Section
10).
9.4.2 Data export
Press the ‘Data export’ softkey to transfer the displayed test record to Audibase (or other computer-based
application). Note that the CA850 needs to be connected to a computer with Audibase correctly configured to
allow this (refer to the Audibase documentation). Data is transferred via the ‘USB PC’ socket.
9.4.3 Print
Press the ‘Print’ softkey to print the audiogram on a printer connected to ‘USB PRINTER’ socket.
Pre-printed
audiogram paper is available for this purpose.
Note: Printer initiation can take several seconds - do not
press the
‘Print’ softkey again until the printout has been generated.