Technical Specifications for Amico GoLift Portable

M

AD

E I

N N

ORTH AM

ER

IC

A

Maximum Lifting Speed:

No load: 1 inch/second
150 lbs/68 kg: 0.6 inches/seconds
450 lbs/204 kg: 0.5 inch/seconds

Unit Weight:

6 lbs

Safe Working Load (SWL):

450 lbs

Dimensions (see diagram):

Length: 19.4" (049.27 cm)"
Height: 5.2" (13.20 cm)
Depth of Lift: 3.5" (8.89 cm)
Strap Length: 74" (187.96 cm)

Number of Lifts per Charge (Duty: 10/90):

89 with 150 lbs (25% of strap at midrange)
52 with 450 lbs (25% of strap at midrange)
Charging time: 2 - 4 Hours

Lift Case:

Flame retardant ABS 
IP27 hand control - 1m water resistance
IP23 Lift - protected from fingers and water 
spray less 60 degrees from vertical

 

Batteries:

High Capacity, Long Lasting NiMH
1 x 24V (1.6Ah)

Battery Charger:

100-240 VAC 40 Watt

Hand Control:

Wireless

Safety:

Emergency Stop: Yes
Emergency Lowering Device: Independent Electrical
Emergency Manual Lowering
Upper Limit Detection
Lower Limit Detection
Slack Tape Sensor
Free Fall Brake (Over Speed Governor)
Low Battery and Dead Battery indication
Soft Start and Stop
Overload Protection

Approvals:

CAN/CSA-C22.2 No 60601-1:08
CAN/CSA-Z10535.1:15

Expected Service Life:

22,500 cycles or 10 years. It is recommended that the lift 
be replaced after this time.
The expected life of slings, batteries fuses, straps and cord 
is dependent upon proper maintenance.

19.4"

18.2"

5.2"

81"

Summary of Contents for Go Lift Portable 450

Page 1: ...Installation and Operation Instructions w w w a m i c o c o m Portable450 Portable450...

Page 2: ...on the Amico Track 10 Using the Reacher Bar 11 Connecting the Hand Control to the GoLift Portable 11 Operating the GoLift Portable 12 Charging the GoLift Portable 12 LED and Buzzer Functions 13 Emerge...

Page 3: ...look under the cover of this compact lift you will find an impressive set of all metal gears and state of the art battery technology that will allow the caregiver to safely and effortlessly transfer a...

Page 4: ...ndards Standard s CAN CSA C22 2 No 60601 1 08 Medical Electrical Equipment Part 1 General requirements for basic safety and essential performance ISO10535 2006 Hoists for the transfer of disables pers...

Page 5: ...checked to ensure that they are in good working order Check for signs of wear or fraying prior to use Report any unusual wear or damage immediately Amico Mobility will not be responsible for any damag...

Page 6: ...ntrol 1m water resistance IP23 Lift protected from fingers and water spray less 60 degrees from vertical Batteries High Capacity Long Lasting NiMH 1 x 24V 1 6Ah Battery Charger 100 240 VAC 40 Watt Han...

Page 7: ...Anatomy of the GoLift Portable Emergency lowering stop feature Unit controls Quick magnetic hand control connection Wireless hand control Durable all metal gears Lifts up to 450 lbs GoLift Portable we...

Page 8: ...hooks of the GoLift Portable Start with the uppermost set of straps from the back and then take the lowest set of straps from the legs Be careful when attaching the lifting sling on the hooks Check t...

Page 9: ...that the patient is sitting up Only persons who have received competent instruction regarding the use of the lifting equipment and fitting of slings should use the Amico GoLift Portable Important Pla...

Page 10: ...arm and place the end cap back on the arm 5 Fasten all end stops tightly using an adjustable wrench NOTE If you are installing the Amico GoLift Portable in an existing track system you must ensure th...

Page 11: ...ssible to hook on 4 Slide the top hook of the Reacher Arm through the trolley Make sure there is enough slack on the lift strap to accomplish this step otherwise it will not be possible to hook on 5 T...

Page 12: ...ake it will override the lift and they will remain paired 2 Press and hold both the buttons on the portable lift for 5 seconds until the lift blinks Green This indicates that the lift is ready to be p...

Page 13: ...connected Not paired Slow flashing Green N A Not connected Paired with main lift Fast flashing Green N A Maintenance Alarm At 1001 lifts the LED will flash Green To reset press Up button then Down but...

Page 14: ...em of the lift malfunctions and the emergency lowering function is used the lift must be reset by a qualified lift technician before re use Manual Emergency Lowering The manual emergency and raising s...

Page 15: ...of the tracks every 4 months using a damp cloth to ensure a better trolling surface for the trolley Troubleshooting Should problems arise with the use of the Amico GoLift Portable review the followin...

Page 16: ...ual noises Inspect the hook components for any signs of damage Inspect the gears for any broken or worn teeth Verify the overspeed cam is operating freely Check emergency stop tab Check emergency lowe...

Page 17: ...Custom Go Track Go Limb Go Turn Go GoLong Polyester SLG GO LONG WV 42X78 SLG GO LONG WV 56X78 GoLong Mesh SLG GO LONG MSH 42X78 SLG GO LONG MSH 56X78 Basic Go Long Go Active Go Comfort Go Hygiene Go C...

Page 18: ...AT NO COPIES ARE TO BE MADE WITHOUT THE CONSENT OF AMICO CLINICAL ALL RIGHTS TO DESIGN OR INVENTION ARE RESERVED DATE DATE CHECKED BY DRAWN BY NTS SCALE 2 FINISH MATERIAL YYMMDD XXX UNLESS SPECIFIED A...

Page 19: ...ed that the lift we brought back to an area kept at room temperature between uses For example after transport in a vehicle do not leave the lift in the trunk or backseat Also try to keep the lift from...

Page 20: ...ishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000...

Page 21: ...IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips shor...

Page 22: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Page 23: ...12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimat...

Page 24: ...months from the time of installation the warranty for power supply is one 1 year from the time of installation This warranty is valid only when the product has been properly installed as outlined in...

Page 25: ...Notes...

Page 26: ...Notes...

Page 27: ...Notes...

Page 28: ...st Beaver Creek Road Richmond Hill ON L4B 1G6 Canada 600 Prime Place Hauppauge NY 11788 USA Toll Free Phone 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 Email amo serv...

Reviews: