background image

12

    Amico Mobility Solutions Corporation

The Amico GoLift Portable is switched on automatically when a button on the 
hand control is pressed. The Amico GoLift Portable switches off automatically 
after approximately three minutes without activation.

To replace the battery in the hand control, use a Phillips screwdriver to remove 
the cover.

To pair & unpair the hand control to the portable lift:

1.  Press and hold both the up and down button simultaneously on the hand control until the LED flashes Red and 

Green (5 seconds), this will put the hand control in paring mode and unpair from the current lift. The hand control 

will search for 10 seconds before going to sleep. If the lift is not found and paired during this time, wake the hand 

control up again by pressing any button. The hand control will wake up still in pairing mode.

NOTE:

 The hand control holds the memory, if the lift is unpaired before the hand control, as soon as the hand control 

is awake, it will override the lift and they will remain paired.

2.  Press and hold both the buttons on the portable lift for 5 seconds, until the lift blinks Green. This indicates that the 

lift is ready to be paired to a new hand control.

Charging the GoLift Portable

The charger plugs in into the port on the GoLift Portable.

The batteries should be charged on a regular basis. It is recommended that the lift be left on charge when not in 
operation, and at the end of each day. This will maximize the life cycle of the batteries.

The GoLift Portable may remain connected to the charger indefinitely since the charger has a built-in regulator, 
eliminating the danger of overcharging.

As a general rule, it is recommended that the lift be raised to a height that will not interfere with anything or
anyone while the lift is not in use. 

!

Do not drain the batteries excessively. This will dramatically reduce the lifespan of the batteries.  
If the slow beep sounds, be sure to recharge the battery as soon as possible.  

!

If the GoLift Portable is not returned to the charger the batteries will be drained and damaged. 

Operating the GoLift Portable

DOWN button

UP button

Summary of Contents for Go Lift Portable 450

Page 1: ...Installation and Operation Instructions w w w a m i c o c o m Portable450 Portable450...

Page 2: ...on the Amico Track 10 Using the Reacher Bar 11 Connecting the Hand Control to the GoLift Portable 11 Operating the GoLift Portable 12 Charging the GoLift Portable 12 LED and Buzzer Functions 13 Emerge...

Page 3: ...look under the cover of this compact lift you will find an impressive set of all metal gears and state of the art battery technology that will allow the caregiver to safely and effortlessly transfer a...

Page 4: ...ndards Standard s CAN CSA C22 2 No 60601 1 08 Medical Electrical Equipment Part 1 General requirements for basic safety and essential performance ISO10535 2006 Hoists for the transfer of disables pers...

Page 5: ...checked to ensure that they are in good working order Check for signs of wear or fraying prior to use Report any unusual wear or damage immediately Amico Mobility will not be responsible for any damag...

Page 6: ...ntrol 1m water resistance IP23 Lift protected from fingers and water spray less 60 degrees from vertical Batteries High Capacity Long Lasting NiMH 1 x 24V 1 6Ah Battery Charger 100 240 VAC 40 Watt Han...

Page 7: ...Anatomy of the GoLift Portable Emergency lowering stop feature Unit controls Quick magnetic hand control connection Wireless hand control Durable all metal gears Lifts up to 450 lbs GoLift Portable we...

Page 8: ...hooks of the GoLift Portable Start with the uppermost set of straps from the back and then take the lowest set of straps from the legs Be careful when attaching the lifting sling on the hooks Check t...

Page 9: ...that the patient is sitting up Only persons who have received competent instruction regarding the use of the lifting equipment and fitting of slings should use the Amico GoLift Portable Important Pla...

Page 10: ...arm and place the end cap back on the arm 5 Fasten all end stops tightly using an adjustable wrench NOTE If you are installing the Amico GoLift Portable in an existing track system you must ensure th...

Page 11: ...ssible to hook on 4 Slide the top hook of the Reacher Arm through the trolley Make sure there is enough slack on the lift strap to accomplish this step otherwise it will not be possible to hook on 5 T...

Page 12: ...ake it will override the lift and they will remain paired 2 Press and hold both the buttons on the portable lift for 5 seconds until the lift blinks Green This indicates that the lift is ready to be p...

Page 13: ...connected Not paired Slow flashing Green N A Not connected Paired with main lift Fast flashing Green N A Maintenance Alarm At 1001 lifts the LED will flash Green To reset press Up button then Down but...

Page 14: ...em of the lift malfunctions and the emergency lowering function is used the lift must be reset by a qualified lift technician before re use Manual Emergency Lowering The manual emergency and raising s...

Page 15: ...of the tracks every 4 months using a damp cloth to ensure a better trolling surface for the trolley Troubleshooting Should problems arise with the use of the Amico GoLift Portable review the followin...

Page 16: ...ual noises Inspect the hook components for any signs of damage Inspect the gears for any broken or worn teeth Verify the overspeed cam is operating freely Check emergency stop tab Check emergency lowe...

Page 17: ...Custom Go Track Go Limb Go Turn Go GoLong Polyester SLG GO LONG WV 42X78 SLG GO LONG WV 56X78 GoLong Mesh SLG GO LONG MSH 42X78 SLG GO LONG MSH 56X78 Basic Go Long Go Active Go Comfort Go Hygiene Go C...

Page 18: ...AT NO COPIES ARE TO BE MADE WITHOUT THE CONSENT OF AMICO CLINICAL ALL RIGHTS TO DESIGN OR INVENTION ARE RESERVED DATE DATE CHECKED BY DRAWN BY NTS SCALE 2 FINISH MATERIAL YYMMDD XXX UNLESS SPECIFIED A...

Page 19: ...ed that the lift we brought back to an area kept at room temperature between uses For example after transport in a vehicle do not leave the lift in the trunk or backseat Also try to keep the lift from...

Page 20: ...ishments other than domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000...

Page 21: ...IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips shor...

Page 22: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Page 23: ...12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimat...

Page 24: ...months from the time of installation the warranty for power supply is one 1 year from the time of installation This warranty is valid only when the product has been properly installed as outlined in...

Page 25: ...Notes...

Page 26: ...Notes...

Page 27: ...Notes...

Page 28: ...st Beaver Creek Road Richmond Hill ON L4B 1G6 Canada 600 Prime Place Hauppauge NY 11788 USA Toll Free Phone 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 764 0862 Email amo serv...