42
Operation
Setup Menu : Supply
3.2.1.2 Min Output Voltage
The minimum output voltage is specified in volts and reflects the
minimum compliance of a connected power supply. The valid range
is 0.000 to -20.000 volts. A unipolar power supply has a minimum
output voltage of 0.000 volts. This setting can be edited only if a
Custom... supply is selected.
3.2.1.3 Max Output Voltage
The maximum output voltage is specified in volts and reflects the
maximum compliance of a connected power supply. The valid range
is 0.001 to +20.000 volts. This setting can be edited only if a
Custom... supply is selected.
3.2.1.4 Min Output Current
The minimum output current is specified in Amperes and reflects
the minimum output current capacity of a connected power supply.
20
-20
200
-200
20
-20
200
-200
V
V
I
I
AMI 12100PS
OPERATING RANGE
AMI 1 MODEL 601
ENERGY ABSORBER
OPERATING RANGE
AMI 4Q05100PS
OPERATING RANGE
5V
Figure 3-3. Example power supply operating ranges.
$ × 0LQ2XWSXW9ROWDJH9
9V
$ × 0D[2XWSXW9ROWDJH9
9V
$ × 0LQ2XWSXW&XUUHQW$
9V