89
Remote Interface Reference
Quench State Control and Queries
4.5.7
Quench State Control and Queries
The
QUench
commands control and query the quench state of the Model
420. For further information regarding the quench detection functions, see
paragraph 3.5 on page 61.
QUench {0|1}
Clears or sets the quenched state. Sending a “0” clears any quench
condition (equivalent to pressing the
5(6(7=(52
front panel switch).
Send a “1” sets a quench condition. Setting the quench state to “1” is
equivalent to a quench detection by the instrument, i.e. the power supply
output is forced to 0 V, the quench output of the rear panel connector J4 is
asserted, and all ramping functions are disabled.
QUench?
Queries the quench state. If a “0” is returned, no quench condition exists.
If a “1” is returned, a quench detect has occurred and is still in effect.
Summary of Contents for 420
Page 2: ......
Page 4: ......
Page 6: ......
Page 12: ...x Rev 7 List of Tables ...
Page 18: ...xvi Rev 7 Foreword Safety Summary ...
Page 22: ...4 Rev 7 Introduction Front Panel Layout ...
Page 24: ...6 Rev 7 Introduction Rear Panel Layout ...
Page 30: ...12 Rev 7 Introduction Operating Characteristics ...
Page 54: ...36 Rev 7 Installation Power Up Procedure ...
Page 82: ...64 Rev 7 Operation Summary of Operational Limits ...