No Output Alarm Indication
Resolution(s)
:
•
Verify that the alarm and alarm delay setpoints are correct
•
Confirm the Alarm Delay or Alarm Hold Off setting is correct
•
Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
•
Verify that the alarms on the Analyzer are properly configured using the
COMMAND
CENTER
Software
(see pages 32 – 35)
'Err' Flashes on the LCD
Resolution(s)
:
•
Look up the Error Code on page 44 and troubleshoot/resolve it
•
If you cannot resolve, contact AMI for further support
Analyzer Refuses to Accept Front Panel Settings
Resolution(s)
:
•
Use the
COMMAND CENTER
Software
to verify that the Security Settings match your
preference
Analyzer Reads Zero
Resolution(s)
:
•
Hook up a tank of Span Gas and confirm the Analyzer responds upscale accurately.
No Voltage or Current Output to Recording Device
Resolution(s)
:
•
Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
Display Pressure Reading Not Correct
Resolution(s)
:
•
Perform the Initiation of the Pressure Sensor Procedure on page 15
45