Setting the Alarm Hold Off
Press the ALARM HOLD OFF button, and the Alarm Hold Number will appear in minutes. Within 3-4
seconds, push either the UP or DOWN ARROW BUTTON to adjust the duration of your ALARM HOLD
OFF. The ALARM HOLD OFF can be engaged from 0 to 120 minutes. The HOLD OFF feature holds-off
both ALARMS and ANALOG OUTPUT.
After the time for setting the ALARM HOLD OFF expires, both Alarms and the Analog Output will revert
to measurement mode.
ADDITIONAL NOTES
:
If you need more time for the setup, simply push the ALARM HOLD OFF Button again, and it will
automatically reset to the original Hold Off Time.
If you are completing a Calibration before the 'Hold Off' Set Time elapses and want the Alarms and
Analog Output to become functional immediately, you can simply run the Hold Off Time to zero by
pushing the Hold Off Button until the LCD blinks and then pushing the DOWN ARROW BUTTON until the
LCD shows zero.
NOTE
:
The ALARM HOLD OFF allows you to bypass the Alarm Relay Function for a predetermined amount of
time. The feature is helpful to use during monthly or quarterly gas calibrations so as not to set off alarm
components driven by the Relay contacts.
Up and Down
Arrow Buttons
Alarm Hold off