PN 9000-280-VE, Rev B
Remote Calibration Unit (RCU) Specifications
Specification
Description
Display
4.2" color 1/4W VGA with graphical user interface,
password-protected.
Keypad
18-key membrane
Input
Two-wire Modbus RTU (19200 Baud Rate, Even Parity, 1 Stop Bit)
from analyzer. Host capable of up to eight analyzers.
Digital Outputs
Two- or four-wire Modbus RTU, TCP/IP Ethernet with embedded
web server (RJ-45 connection), USB port for data collection or
software update.
Environment
Ambient Temperature:
-18 °C to 60 °C (0 °F to 140 °F)
Maximum Altitude:
2000 meters
Relative Humidity:
0 to 90 %, non-condensing
Enclosure
UL Type 4X (NEMA 4X (IP66), Division 2
Power Requirements
Nominal 115 to 230 VAC ±10 %, 47–63 Hz, 75 VA maximum
Specifications
|
4-3