Versitron Operation Manual
2.1
ELECTRICAL SAFETY
1. The Versitron Test Machine has been designed to meet the requirements of BS
EN 61010-1:2010 Safety requirements for electrical equipment for measurement,
control and laboratory use.
2. The user manual contains some information and warnings, which have to be
followed by the user to ensure the safe operation and to keep the machine in safe
condition.
3. The machine has been designed for indoor use.
4. This machine complies with electrical safety grade Class 1, which means that it is
“earthed apparatus” and requires the mains plug to contain a protective earth
terminal. The mains plug must only be inserted in a socket outlet provided with a
protective earth contact. The protective action must not be negated by the use of
an extension cord without a protective conductor.
5. Make sure that only fuses with the required rated current and of the specified type
are used for replacement, see Section 7.5. The use of makeshift fuses and the
short-circuiting of fuse holders is prohibited.
6. There are no user serviceable parts within the machine.
7. The machine must be disconnected from all voltage sources before it is opened
for any adjustments, replacement, maintenance or repair.
8. Capacitors inside the machine may still be charged even if the machine has been
disconnected from all voltage sources.
9. Any adjustment, maintenance or repair of the opened machine connected to the
mains supply voltage should be avoided as far as possible but if inevitable, it must
only be carried out by a skilled person who is aware of the hazards involved.
10. Only use safety approved power cords that come with the Versitron test
machines.