34 | Model ta7000F Gas Purity Monitor
After pressing Warmup the instrument will begin heating critical zones
and will perform a series of system tests. The results from these tests will
appear on the Warmup screen. The Warmup menu item will also change
to Cool. If you wish to stop the warm-up process, press the Cool key and
the instrument will return to a completely neutral state. If any errors oc-
cur during the warm-up process, proceed to Error messages on page 46.
Warmup Mode Screen
The Warm-up process typically takes 20 to 30 minutes. When the warm-
up process has completed succesfully a
Run
munu label will appear.
Pressing the
Run
key will take the user to the main screen where the Main
menu will be available.
Warmup Pass Screen
Warmup Mode
12:35:25
Page
Module Electronics: Pass
Detector System:
Detector Temperatures:
Carrier Pressure:
Cool
WARM-UP
WARM-UP
Idle Mode
12:34:25
Module Electronics:
Detector System:
Detector Temperatures:
Carrier Pressure:
Pass
Pass
Pass
Pass
***** Successful Warm-up *****
Run Cool
Page