TR-2100 User Manual
30
Cross Triggering
Any number of TR-2100 recorders may be coupled together to form a system with any
number of inputs (in multiples of 8, 16, 24, or 32 analogue inputs). This system uses the
'Triggered' status relay outputs and digital input 15 of each recorder. They are connected as
follows:
C
15
3
C
15
3
C
15
3
C
15
3
B
Battery -
Digital inputs
Triggered
outputs
Figure 7 Cross-Triggering Wiring
Switch 5 in bank SW1 on the acquisition board must be ON for every machine connected
together in this way. When this is done and the systems are switched on the channel labels
for digital input 15 will become 'Cross-trigger'. The triggering for these channels must be
enabled manually but are automatically set as level sensitive with profiling off.
Operation
When any of the connected recorders is triggered, its 'Triggered' status output relay will
pulse immediately. This will cause digital input 15 on each recorder to go into the alarm
state. When this happens, the recorders will trigger, but their status relays will not be
activated. If a valid trigger is detected on a recorder other than that which started the
process, its status relay will activate, thus extending the post fault period of all recorders if
necessary.
Summary of Contents for TR-2100
Page 2: ......
Page 6: ...iv ...
Page 10: ...TR 2100 User Manual 4 ...
Page 16: ...TR 2100 User Manual 10 ...
Page 24: ...TR 2100 User Manual 18 ...
Page 34: ...TR 2100 User Manual 28 ...
Page 38: ...TR 2100 User Manual 32 ...
Page 54: ...TR 2100 User Manual 48 Figure 12 Analogue Input Board Jumpers ...
Page 78: ...TR 2100 User Manual 72 ...
Page 94: ...TR 2100 User Manual 88 ...
Page 98: ...TR 2100 User Manual 92 ...
Page 99: ...TR 2100 User Manual 93 Appendix V System Drawings ...
Page 100: ...TR 2100 User Manual 94 ...
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......