Controller / User Interface | 4-31
Alarm3= Cal/Verif
y, depending on whether alarm 3 should be
associated with % oxygen levels, or with the start of a calibration or
verification, or with the start of a calibration or verification (dual
function).
The control unit automatically exits you from this menu after you make
your selection.
Service Alarm
The service alarm operates as normally closed (service alarm
condition causes open contacts).
The service alarm provides an alarm that alerts you of possible system
problems that require attention. Events that can trigger the service alarm
to activate are as follows:
y
Error condition (detected by software)
y
System operational variance (such as an over-temperature or under-
temperature condition)
y
Calibration failure
y
Verification failure
y
Control unit power failure
Watchdog Relay
During normal system operation, the watchdog relay operates as
normally closed (system failure causes open contacts).
The watchdog alarm relay de-energizes when the system watchdog timer
times out. The watchdog timer will time out and cause the control unit
to reset itself to prevent the control unit from becoming “lost” due to
such things as noise spikes or power surges. You can use this relay to
activate a remote indicator to notify you if the watchdog timer times out.
NOTE
NOTE
Summary of Contents for Thermox 2000 ATEX Series
Page 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Page 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Page 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Page 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Page 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Page 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...