5-14 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
9. Press the Enter key once the combustibles span gas reading(s) stabi-
lize on the far left of the display.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Calibrating
The system then returns to monitoring process readings.
Be sure to reinsert the 1/4" tube cap on the calibration inlet port on
the sensor once you have completed the manual calibration. This
prevents air from getting into the sample gas line during normal
operations.