4-30 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Relay Configure
All Series 2000 control unit relays are Form A, normally open contacts.
You can, however, use the software to set whether your alarms should
close on an alarm condition (energize on alarm) or open on an alarm
condition (de-energize on alarm). This menu option applies to all alarms
except for the service and watchdog alarms, which are always set to de-
energize on alarm (fail-safe).
1. Select
Relay Configure
from the Alarm menu.
2. Select
De-energize on Alarm or Energize on Alar
m.
If you select Energize on Alarm, the following message will appear:
Relays are set to
ENERGIZE on Alarm
If you select De-energize on Alarm, the following message will appear:
Relays are set to
DE-ENERGIZE on Alarm
Select Function
This menu option allows you to select which function alarm 3 should
track:
y
% Oxygen
y
When the system starts a calibration
y
When the system starts a verification
y
When the system starts a calibration
or
a verification (dual function)
If tracking the start of a calibration or verification, the alarm will deacti-
vate once the calibration or verification has been completed, including
any recovery time you have specified (see the Calibrate Key Chapter for
further details on calibrations and verifications).
1. Choose
Select Function
from the Alarm menu.
2. Select
Alarm3= Oxyge
n,
Alarm3= In Cal
.,
Alarm3= In Verify
, or
Summary of Contents for Thermox 2000 ATEX Series
Page 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Page 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Page 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Page 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Page 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Page 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...