Controller / User Interface | 4-7
Display
This menu option allows you to define what information should appear
on the top three display lines of your Series 2000 control unit. Choices
for each line are as follows:
y
Percent oxygen (% O
2
)
y
Cell temperature
y
Cell millivolts
y
Thermocouple millivolts (T/C millivolts)
y
Current date/time
y
User text
y
Turn off (leave display line blank)
y
Combustibles (optional)
When you access menus, menu options can occupy display lines 3
and 4, and sometimes display line 2. Therefore, we recommend that
you place the information that is most critical for you to view on
the topmost display lines.
Setting up the display
1. Select
Display
from the Setup menu.
2. Select
Display Line #
1,
Display Line #
2, or
Display Line #
3,
depending on where you want to place the information.
3. Select the type of information you want to place on the display line.
Entering User Text
The Down arrow appearing on the third line of the display allows you to
see your current position when entering your user-text message. When
you first select the User Text option, this Down arrow will appear above
the first position.
To enter text, use the Up and Down arrow keys on your control unit
keypad to scroll through the ASCII text characters. Once the letter you
want to enter appears in that position, press the Right arrow key to
move to the next position. You can also press the Left arrow key to move
back one position in your text message. When you are finished entering
your message, press the Enter key.
NOTE
Summary of Contents for Thermox 2000 ATEX Series
Page 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Page 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Page 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Page 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Page 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Page 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Page 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...