28
12 COUPLING MODE SELECTION AT THE GENERATOR USER INTERFACE
The possible selections are:
Surge output
Manual CDN
IEC coupling
12.1 Surge Output
If a CDN 3153-S63.1 3-phase coupler is connected to the generator, this CDN will always be
taken as default unit and the 3-phase coupling modes is shown at the user interface.
The selection “Surge Output” is for using the HV pulse output plugs for any surge application
without any CDN or other selectable device. The surge output selection will activate the HV
pulse output. This can be a verification or an application to external equipment (no Teseq
product).
No coupling to the EUT lines of the connected CDN will be realized.
12.2 Manual CDN
This selection must be used if an external manual coupler is connected; like CDN 3083 for
surge pulse. With this selection, the loss of an external CDN will be compensated.