16
9.2 Rear panel CDN 3153-S63.1
9.2.1 Mains input and power switch
The main input is the power to the internal electronics of the CDN.
This connection must not be confused with the EUT power input.
The input contains the mains power input connector, the mains switch and the mains fuses.
Before putting the instrument into operation, please make sure that the
voltage range shown on the mains input module corresponds with the
voltage of the local supply to which the instrument will be connected, and
whether the fuses are correctly rated (2 x 3.15 AT).
The used cable must be rated for a current of at least 3.5 A