Operation
SW 5550A•SW 3700A•SW 1850A
3-24
Operation Manual
SYNC OUT
selects the output sync signal to ON, OFF or EVENT. In the
ON
position,
the signal is driven by the output frequency of the waveform, or sequence settings. In
the
EVENT
position, a pulse will be generated for any change in VOLT, CURR, FREQ,
PHASE or WAVEFORM.
The
SYNC OUT
setting only affects normal (non-sequence) mode. To set up the SW
so that a sync pulse only occurs for a specified sequence segment, set this parameter
to OFF. That way, before and after a sequence is run, no other sync pulses will occur.
Table 3–2 provides self-test result definitions.
BIT
MEANING
BIT
MEANING
0001
LCD Test
0100
Battery Test
0002
EPROM CKSUM
0200
QSPI Test
0004
NVRAM BANK 0 CKSUM
0400
DFI Test
0008
NVRAM BANK 1 CKSUM
0800
Speaker Test
0010
RAM Test
1000
DWSB Self-Test 0
0020
VXI Test
2000
DWSB Self-Test 1
0040
QSCI Test
4000
T&MB Self-Test 0
0080
GPIB Test
8000
T&MB Self-Test 1
Table 3–2 SELF-TEST Results Definitions