SW 5550A•SW 3700A•SW 1850A
Operation
Operation Manual
3-37
3.5.2
Voltage Spike Example
This example illustrates programming the SW for a voltage spike on a 150V sine
waveform that starts at 30° and ends at 50° with an amplitude of 220.6 volts.
PROGRAM
MEASURE
WAVEFORM
SEQUENCE
INSTR
SYSTEM
MAIN MENU
OUTPUT [OFF]
1.
At the MAIN MENU, select the
WAVEFORM
menu, then press <
Enter
>.
LOAD
EDIT
VIEW
SAVE
DELETE
WAVEFORM
OUTPUT [OFF]
2.
The WAVEFORM Menu will default to the LOAD sub-menu. If the cursor is not
pointing to LOAD, use the Knob or Up/Down Arrow keys to select
LOAD
.
3.
Press the
Enter
key to enter the
FUNC
(function) mode.
4.
Press <
Enter
> again to place a box around the waveshape selection.
5.
Use the Knob to cycle through the waveshape selections until
SINE
appears in
the FUNC box.