Cleaning Precautions

The  sealant  used  between  the  prism  and  stainless  steel  well  may  be  attacked  by  certain 

chemicals. Reichert recommends avoiding strong solvents such as N,N-Dimethylformamide or 

N,N-Dimethylacetamide, Tetrahydrofuran, cresols, phenols, and other tar acids. Some lacquer 

thinners can also deteriorate the seal over time. To determine if your samples or solvents may 

attack this seal, please contact Reichert Analytical technical support.

Scratched optical surfaces can have an adverse effect on optical performance. In the case of 

a sample having been allowed to dry on the prism surface, never use abrasive materials to 

remove that sample. Instead redissolve the sample with an appropriate solvent to protect the 

prism surface from being scratched. Thick, sticky samples should be rinsed off with a suitable 

solvent to avoid possibly scratching the prism. Use caution when removing samples or drying 

the prism. 

Care and Cleaning

Preparing the prism and well surfaces for samples is essential if consistent 

and accurate values are to be attained. Prior to and immediately after a 

sample is placed on the prism surface, those surfaces must be properly 

cleaned (Fig 4). It is imperative that these surfaces be cleaned to assure 

removal of any possible residue since this condition will have an adverse 

affect on resultant readings. Never allow a sample to dry on the prism or well 

surfaces. 

Depending upon the sample, clean the prism and well using distilled water, diluted soapy water, 

isopropyl alcohol or methanol. Rinse with distilled water and gently dry using a lint free product 

like Kim Wipes® that will leave no residue.

To properly clean the body or display areas of the instrument simply wipe with a soft cloth and a 

mild soap and water solution. A typical house hold glass cleaning product works well.

Fig 4

11

Summary of Contents for Reichert r2i300

Page 1: ...i300 REFRACTOMETER USER GUIDE...

Page 2: ......

Page 3: ...recommends that daily calibration with distilled water be performed as a minimum requirement To assure the longest life of your r2 i300 always store the instrument with the prism surfaces and instrume...

Page 4: ...n 6 Set Point Calibration 6 Multi Point Calibration 6 Add Delete Edit Calibration Points 7 Operational Parameters Sample Temperature Effects on Readings 7 Automatic Temperature Correction TC 7 Compute...

Page 5: ...power input connector or the unit s power cord Damage to the r2i300 instrument and or injury to the operator may occur Warning The r2i300 should be used in strict accordance with the instructions outl...

Page 6: ...Voltage 100 240VAC 47 63HZ 1 35A Power Output 40W max 5 VDC 8A Fuse Rating 500mA 250V 5 x 20mm quick acting Power Cords US EU UK supplied Dimensions 12 L x 8 75 W x 3 25 H 30 9cm x 22 5cm x 8 4cm Net...

Page 7: ...Intrument Features Calibration Oil Sample Well Power Input Connector Power Switch Fuse Holder Com1 PC DB9 Female Com2 Printer DB9 Male Back Panel 4 Touch Screen Display Menu Key Read Key Calibrate Key...

Page 8: ...e on position for a minimum of one hour before attempting calibration or to measure samples Fig 1 5 Sample Analysis Cleanliness sample quality and sample temperature equilibration are essential to obt...

Page 9: ...Water appears press and follow the displayed instructions to Apply distilled water Allow adequate time for the sample to temperature equilibrate and press Clean and dry the sample surface and well an...

Page 10: ...iewing instrument readings One standard method is to have programmed custom channels that calculate the correct reading at 20 degrees C for a sample that is being read at other temperatures The use of...

Page 11: ...of the instrument through the Menu Options and activate that option To complete the link between instrument and computer click on the Connect PC Comm pick found in the open software window If success...

Page 12: ...he supplied PC interface software The following describes the r2 i300 custom channel possibilities Channel 1 ICUMSA BRIX The r2 i300 is pre programmed at the factory with channel 1 to read Sucrose or...

Page 13: ...Reading Log is displayed and is pressed the entire reading log will be printed 1 to 1000 readings Delete Reading Log Will delete stored values when is pressed Calibration Covered in greater detail und...

Page 14: ...suitable solvent to avoid possibly scratching the prism Use caution when removing samples or drying the prism Care and Cleaning Preparing the prism and well surfaces for samples is essential if consis...

Page 15: ...s in the refractometer may be required Warranty and Non Warranty Service The warranty period for this instrument is two years against manufacturing defects Please visit our web site at www reichertai...

Page 16: ...8955 Phone 716 686 4500 email reichertai refractometers ametek com www reichertai com AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Tel 49 89 315...

Reviews: