background image

Sample Temperature Effects on Readings

Refractive Index is a physical property of a material. Refractive index measurements, and the 

percent solids results derived from them, are very dependent upon temperature. For most 

samples, raising the temperature will lower the refractive index of a sample. Conversely, 

lowering the sample temperature will raise the refractive index. This change in refractive index, 

due to a change in temperature, will cause an apparent change in concentration unless the 

reading is temperature corrected (TC). Before a sample measurement can be temperature 

corrected however, the sample needs to reach thermal equilibrium with the instrument. Without 

temperature equilibration, the resulting value cannot be considered accurate for direct read 

values or temperature corrected (TC) values. This is the result of the sample being in a state of 

thermal transition while the refractive index value is captured by the instrument. This is common 

with all refractometers. 

Automatic Temperature Correction (TC)

Automatic Temperature Correction (TC) is based upon three factors, refractive index, 

temperature and temperature coefficient.  The relationship of these three factors are used 

to mathematically correct a value as though it was read at a specific temperature. The most 

common temperature corrected (TC) temperature is 20 degrees C but it can differ with 

individual requirements. The temperature corrected reading modes (TC) in this instrument, 

are based upon sucrose temperature correction factors developed by ICUMSA. Non-sucrose 

solutions will have unique temperature correction factor.

Since most uses of a refractometer is to control concentration, it is useful to eliminate the 

temperature effects when viewing instrument readings. One standard method is to have 

programmed custom channels that calculate the correct reading at 20 degrees C for a sample 

that is being read at other temperatures. The use of a custom channel allows comparing 

readings to a formula or process specification at a predetermined temperature without doing 

complicated conversions.  Please note, that custom channels to do temperature correction, are 

specific to the material that the channel was developed for, and most likely will provide poor 

readings if used for a different material.

7

Add/Delete/Edit Calibration Points

Calibration fluid values may change from one lot to the next. It is also possible to use alternate

fluids to calibrate this instrument. As a result, it will become necessary to edit the stored values

to match any new calibration standard through the “Calibration” option menu. 

To access these options from the Calibration menu press            

to advance to the Add/Delete/Edit option. Press            to accept 

and enter the following option choices.

Add

 - To add a point or turn a stored calibration point into an 

active point advance to “Add” and press           . Advance through 

the available options by pressing           . Once the optional point 

is identified press            to accept. 

Delete

 - To delete a point or turn a calibration point into an 

inactive point advance to “Delete” and press           . Advance 

through the available options by pressing           . Once the 

optional point is identified press            to accept. 

Edit 

- To edit a point press           until desired point is displayed 

and press            to accept. Change values by pressing            or      

           to increase or decrease the value of the indicated (^)  

numeral. Once the value is correct press            menu to accept. 

Repeat this process until all indicated (^) values match that of the 

standard that will be used for calibration.

Add/Delete/Edit

 

Brix-TC                              Brix

Edit

 

Brix-TC                              Brix

Delete

 

Brix-TC                              Brix

Add

 

Brix-TC                              Brix

Read

Read

Read

Read

Read

Menu

Menu

Menu

Menu

Menu

Menu

Menu

Calibrate

Summary of Contents for Reichert r2i300

Page 1: ...i300 REFRACTOMETER USER GUIDE...

Page 2: ......

Page 3: ...recommends that daily calibration with distilled water be performed as a minimum requirement To assure the longest life of your r2 i300 always store the instrument with the prism surfaces and instrume...

Page 4: ...n 6 Set Point Calibration 6 Multi Point Calibration 6 Add Delete Edit Calibration Points 7 Operational Parameters Sample Temperature Effects on Readings 7 Automatic Temperature Correction TC 7 Compute...

Page 5: ...power input connector or the unit s power cord Damage to the r2i300 instrument and or injury to the operator may occur Warning The r2i300 should be used in strict accordance with the instructions outl...

Page 6: ...Voltage 100 240VAC 47 63HZ 1 35A Power Output 40W max 5 VDC 8A Fuse Rating 500mA 250V 5 x 20mm quick acting Power Cords US EU UK supplied Dimensions 12 L x 8 75 W x 3 25 H 30 9cm x 22 5cm x 8 4cm Net...

Page 7: ...Intrument Features Calibration Oil Sample Well Power Input Connector Power Switch Fuse Holder Com1 PC DB9 Female Com2 Printer DB9 Male Back Panel 4 Touch Screen Display Menu Key Read Key Calibrate Key...

Page 8: ...e on position for a minimum of one hour before attempting calibration or to measure samples Fig 1 5 Sample Analysis Cleanliness sample quality and sample temperature equilibration are essential to obt...

Page 9: ...Water appears press and follow the displayed instructions to Apply distilled water Allow adequate time for the sample to temperature equilibrate and press Clean and dry the sample surface and well an...

Page 10: ...iewing instrument readings One standard method is to have programmed custom channels that calculate the correct reading at 20 degrees C for a sample that is being read at other temperatures The use of...

Page 11: ...of the instrument through the Menu Options and activate that option To complete the link between instrument and computer click on the Connect PC Comm pick found in the open software window If success...

Page 12: ...he supplied PC interface software The following describes the r2 i300 custom channel possibilities Channel 1 ICUMSA BRIX The r2 i300 is pre programmed at the factory with channel 1 to read Sucrose or...

Page 13: ...Reading Log is displayed and is pressed the entire reading log will be printed 1 to 1000 readings Delete Reading Log Will delete stored values when is pressed Calibration Covered in greater detail und...

Page 14: ...suitable solvent to avoid possibly scratching the prism Use caution when removing samples or drying the prism Care and Cleaning Preparing the prism and well surfaces for samples is essential if consis...

Page 15: ...s in the refractometer may be required Warranty and Non Warranty Service The warranty period for this instrument is two years against manufacturing defects Please visit our web site at www reichertai...

Page 16: ...8955 Phone 716 686 4500 email reichertai refractometers ametek com www reichertai com AMETEK GmbH Business Unit Reichert Carl von Linde Strasse 42 85716 Unterschleissheim Munich Germany Tel 49 89 315...

Reviews: