Guidance & Manufacturer’s
Declarations
(contin-
ued)
Table 202 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Equipment and Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ClearChart 4P is suitable for use in all establishments and is intended for use in the
electromagnetic environment specified below. The customer or user of the ClearChart 4P
should ensure that it is used in such an environment.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
Power frequency magnetic fields
should be that of a typical residential,
commercial or hospital environment.
30 A/m
30 A/m
Electromagnetic
Environment
- Guidance
Immunity
Test
IEC 60601
Test Level
EFT
IEC 61000-4-4
Mains power quality should be that
of a typical residential, commercial or
hospital environment.
±0.5kV, ±1kV,
±2kV Mains
±0.5kV, ±1kV I/Os
±0.5kV, ±1kV,
±2kV Mains
±0.5kV, ±1kV I/Os
ESD
IEC 61000-4-2
Floors should be wood, concrete or
ceramic tile. If floors are synthetic,
the r/h should be at least 30%.
±2kV, ±4kV,
±8kV Contact
±2kV, ±4kV,
±8kV, ±15kV Air
±2kV, ±4kV,
±8kV Contact
±2kV, ±4kV,
±8kV, ±15kV Air
Compliance
Level
Surge
IEC 61000-4-5
Mains power quality should be that
of a typical residential, commercial or
hospital environment.
±0.5kV, ±1kV
Differential
±0.5kV, ±1kV,
±2kV Common
±0.5kV, ±1kV
Differential
±0.5kV, ±1kV,
±2kV Common
Mains power quality should be that
of a typical residential, commercial or
hospital environment. If the user of the
ClearChart 4P requires continued opera-
tion during power mains interruptions,
it is recommended that the ClearChart
4P be powered from an uninterruptible
power supply or battery.
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
100% Dip
for 0.5 Cycle
at 0°, 45°, 90°,
135°, 180°, 225°,
270°, and 315°
100% Dip
for 1 Cycle at 0°
Voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
100% Dip
for 0.5 Cycle
at 0°, 45°, 90°,
135°, 180°, 225°,
270°, and 315°
100% Dip
for 1 Cycle at 0°
13785-101 Rev. F