16060-101 Rev. M
24
Instructions For Use
(
continued)
Alignment & Measurement
(continued)
Measurement Count
The eye icons below the eye titles (Right and Left, or OD and OS) will be filled in to indicate the
number of measurements that have been made.
When using the single measurement function, an eye icon will become filled after each
measurement. If more than 3 individual measurements are made on the same eye, all three eye
icons will remain filled and the measurement results will continue to be updated based on the
averaging methodology described in the Measurement Results section of this manual.
When using the multi measure function, all eye icons will appear filled after the measurement
process is complete.
Pressing the multi measure icon again will result in a new set of measurements, replacing any
data displayed for the eye being measured.