16060-101 Rev. M
9
Instrument Setup
(
continued)
Application of Input Power
WARNING:
CARE MUST BE TAKEN TO ARRANGE THE CABLES FOR THE ACCESSORIES SUCH THAT THEY
DO NOT PRESENT A TRIPPING HAZARD TO THE EXAMINER OR A DANGER TO THE PATIENT.
WARNING:
POSITION THIS INSTRUMENT SO THAT IT IS NOT DIFFICULT TO OPERATE THE DISCONNECTION
DEVICE (PLUG).
1. After the unit is in its secure location, apply the correct input voltage to the instrument using the
Power Cord from the Accessory Tray.
2. Press down on the “|” located on the ON/OFF Switch. The power inlet is located on the
underside of the unit (Refer to page 10, item 8, for its location).
3. Read the User’s Guide and the Quick Reference Card before operating this instrument.
WARNING:
DO NOT REMOVE THE OUTSIDE COVERS OF THE UNIT OR ATTEMPT TO REPAIR ANY INTERNAL
PARTS. REPAIR AND SERVICE OF THE UNIT MUST BE PERFORMED BY EXPERIENCED PERSONNEL OR
DEALERS THAT ARE TRAINED BY REICHERT.
CAUTION:
ENSURE THAT THE VOLTAGE APPLIED TO THE UNIT IS THE SAME AS THE VOLTAGE THAT IS
INDICATED ON THE DATA PLATE NEXT TO THE INPUT CORD RECEPTACLE OR DAMAGE TO THE UNIT MAY
OCCUR.
CAUTION:
FOR CONTINUED PROTECTION AGAINST THE RISK OF FIRE, ANY REPLACEMENT OF DAMAGED
FUSES MUST BE IN ACCORDANCE WITH THE RATING AS INDICATED IN THE SPECIFICATIONS SECTION OF
THIS MANUAL.
Disconnection of Input Power
1. At any time, the power switch can be set to OFF. The unit does not have a power down sequence.
To terminate operation of this instrument, press the ON / OFF switch to the OFF position (O).
2. If this instrument is intended to be OFF for an extended period of time, it can be disconnected from
power by detaching the power cord from the its receptacle.