ENGLISH 4/12
Follow the instructions below before using the calibrator (cf. Fig. 1):
1.
Place the calibrator on an even horizontal surface away from
all draughts.
Caution…
Do not
use the instrument if the ventilator is out of order.
Ensure a free supply of air to the ventilator located at the
bottom of the instrument (pos. 9).
2.
Check that the voltage shown on the power control switch
(pos. 7) is identical to the mains voltage.
3.
Plug in the cable below the power control switch (pos. 8) and
check that the earth connection is present.
4.
Select an insertion tube (pos. 4) with a well diameter that
matches the sensor (pos. 1) to be calibrated. Insert the
insertion tube in the well of the calibrator (note that the well
and the insertion tube
must
be clean before use).
5.
If the design of the sensor permits it, use an insulation tube
(pos. 2) and insulation (pos. 3) for the best results. These
should be mounted as shown in Fig. 1.
6.
Place the sensor (pos. 1) in the insertion tube (pos. 4) as
shown in Fig. 1.
7.
In order to spare the sensor and its connections it is
recommended to use a heat protection shield (104216) at
high temperatures.
Summary of Contents for JOFRA 320 A
Page 2: ......
Page 4: ...123199 01 11 09 2003 AUTO STEP SWITCH TEST ESC MENU min C F F C CALIBRATION INSTRUMENTS FIG 2...
Page 5: ...11 09 2003 123199 01 C C...
Page 6: ...123199 01 11 09 2003...
Page 7: ...11 09 2003 123199 01...
Page 8: ...123199 01 11 09 2003...
Page 10: ......
Page 36: ......
Page 50: ......
Page 64: ......
Page 78: ......
Page 92: ......
Page 93: ...1 14 1 0 2 1 1 2 2 0 3 2 1 3 2 2 5 2 3 6 2 4 7 2 5 7 2 6 7 2 7 8 2 8 9 2 9 10 3 0 11 4 0 13...
Page 94: ...2 14 1 0 Artvik AMETEK Denmark A S 1 1 1 1 1 1 1 1 1 1 1 1 1 RS232 1 AmeCal Temperature Light...
Page 95: ...3 14 2 0 2 1 1 10 11 50 20 1 CTC 1200A...
Page 96: ...4 14 1 1 9 2 7 3 8 4 4 1 5 2 3 1 6 1 4 1 7 104216...
Page 97: ...5 14 2 2 2 ENTER ESC MENU 2 AUTO STEP AUTO STEP SWITCH TEST SWITCH TEST...
Page 98: ...6 14 2 3 2 F F AUTO STEP...
Page 99: ...7 14 2 4 1 IEC950 2 5 3 2 6...
Page 100: ...8 14 2 7 SWITCH TEST 4 Tmin Tmax Tmin Tmin Tmax...
Page 101: ...9 14 2 8 AUTO STEP 5...
Page 102: ...10 14 2 9 MENU 6 SETUP 2 SETUP C F SWITCH TEST 0 1 SETUP...
Page 103: ...11 14 CA L IB R AT I O N IN S T RU ME NTS 3 0 230V 115V...
Page 105: ...13 14 4 0 100 C 1 1 100 C 100 C 2 0 C 50 C 3 1 7...
Page 106: ...14 14 4 5...
Page 109: ...2 0 2 1 1 10 11 50V 20cm 1m Ametek Denmark A S CTC 1200 A 3 14 JAPANESE...
Page 110: ...1 1 9 2 7 3 8 4 1 4 5 2 3 1 6 1 1 4 7 104216 JAPANESE 4 14...
Page 111: ...5 14 JAPANESE 2 2 2 ENTER ESC MENU 2 AUTO STEP AUTO STEP SWITCH TEST SWITCH TEST...
Page 112: ...2 3 2 F F AUTO STEP AUTO STEP SWITCH TEST SWITCH TEST JAPANESE 6 14...
Page 113: ...2 4 1 RS232 PC IEC950 2 5 3 2 6 7 14 JAPANESE...
Page 114: ...2 7 SWITCH TEST SWITCH TEST 4 Tmin Tmax Tmin Tmin Tmax 3 JAPANESE 8 14...
Page 115: ...2 8 AUTO STEP AUTO STEP 5 9 14 JAPANESE...
Page 116: ...2 9 MENU 6 SETUP 2 SETUP SETUP C F SWITCH TEST 1 0 1 SETUP SETUP 2 JAPANESE 10 14...
Page 117: ...3 0 2 230V 115V 2 11 14 JAPANESE...
Page 119: ...4 0 100 C 212 F 1 1 100 C 212 F 100 C 212 F 2 0 C 32 F 50 C 122 F C 140 3 7 13 14 JAPANESE...
Page 120: ...JAPANESE 14 14 4 5...
Page 122: ...1 0 CTC 1 1 1 1 1 1 1 1 1 1 1 1 RS232 AmeCal Temperature Light CHINESE 2 13...
Page 123: ...2 0 2 1 1 10 11 50V 20 cm 1 m Ametek Denmark A S CTC 1200 A 3 13 CHINESE...
Page 124: ...1 1 9 2 7 3 8 4 1 4 5 2 3 1 6 1 4 1 7 104216 CHINESE 4 13...
Page 126: ...AUTO STEP SWITCH TEST SWITCH TEST 2 4 1 RS232 PC IEC950 2 5 3 2 6 CHINESE 6 13...
Page 127: ...7 13 CHINESE...
Page 128: ...2 7 SWITCH TEST 4 Tmin Tmax Tmin Tmin Tmax CHINESE 8 13...
Page 129: ...2 8 AUTO STEP 5 9 13 CHINESE...
Page 130: ...2 9 MENU 6 SETUP 2 SETUP C F SWITCH TEST 0 1 SETUP CHINESE 10 13...
Page 132: ...CTC 1200 115V 6 3AT 60B313 230V 3 15AT 60B312 CHINESE 12 13...
Page 134: ......