To enter values for a new reference sensor
Click a radio button to select sensor 1 or sensor 2. And
press “Upload” to read current coefficients.
Type in a unique and descriptive name for the reference
sensor.
Use the mouse to position the pointer in the boxes in the
table.
Select “ITS-90” or “Callendar van Dusen coefficients” and
type in the coefficients from the Calibration Certificate
supplied with the reference sensor.
When all the values from the Calibration Certificate have
been entered, click “Download” values.
When the values are downloaded they overwrite the ones
stored in the indicator.
Note…
These coefficients are not saved. Therefore, they must be
typed in and downloaded when they need to be used.
8.2.4
Default values
This option downloads a set of default values, which are stored in the
AmeTrim-ATC/DTI software.
Press “Yes” to download default values to the DTI.