pressing on the drop down button underneath it will display a second drop down list.
This list provides test results that are based on the force tester measurements and contains
"NONE", "PEAK LOAD", "BREAK LOAD", "AVERAGE LOAD", "LOAD AT LIMIT",
"DISTANCE AT PEAK LOAD", "DISTANCE AT BREAK", "DISTANCE AT PEAK" and
"DISTANCE AT LIMIT".
The selected result will be available for printing and exporting (see later) but can also be
displayed on the graph screen after each test. To display this result, check the checkbox to
the left of the result selectors.
Note that only 5 results can be displayed on the graph screen and any remaining
checkboxes will be disabled after 5 results have been selected.
If a "
TENSION STAGE
" result is selected, the controls to the right hand side of the result
selectors are now enabled and these can be used to provide Pass/Fail criteria for the
selected test result. The Pass/Fail values can be either Min/Max or Variance. Select the
required option then enter the required values and units.
The test will be identified as a "Pass" if
ALL
"Pass/Fail" parameters are met otherwise it will
be identified as a "Fail". The "Pass/Fail" status is shown by the color of the test number
shown at the left hand side of the "Test Screen" as shown in section 11.
If a Pass parameter is a "Peak Load" value between (say) 12 lbf and 18 lbf, then the MIN
value should be set to 12 lbf and the MAX value should be set to 18 lbf.
If any numeric field needs to be edited after an initial value has been entered, press
two times (double tap) on the relevant entry area.
To enter a barcode into the required field, press on the edit box field above the
keyboard to enable the field then scan the barcode.
Note that BOTH values must be entered, so if a Pass parameter is a "Break Load" value of
greater than (say) 12 lbf, then the MIN value should be set to 12 lbf and the MAX value
should be set to the load cell capacity, e.g. 225 lbf.
CS User Manual
57
Summary of Contents for Chatillon CS1100
Page 2: ...Part No 01 3953 MAY 2015 CS225 CS1100 Force Tester User Manual...
Page 50: ...Please refer to section 10 for details regarding test samples CS User Manual 49...
Page 126: ...CS User Manual 125...
Page 127: ...21 0 SPARE PARTS KITS SPK CS225 Sheet A 126 CS User Manual...
Page 128: ...SPK CS225 Sheet B CS User Manual 127...
Page 131: ...SPK CS1100 Sheet A 130 CS User Manual...