17.3 ASCII COMMANDS
There are many ASCII commands that can be used to control or read data from the Force
Tester and the next section outlines the commonly used ones.
A complete list of ASCII commands is shown in section 17.5.
All commands require a single alphabetical character. However, some commands require
additional numerical information and this is indicated by
???
in the text below.
Note that these values are entered without units but the units used are those specified as
the "Default" units set in the Force Tester configuration screen.
For example, if the "Default" LOAD unit is pounds force (lbf) then setting the Upper LOAD
limit to
H100.0
(see later) specifies a LOAD limit of 100lbf.
Every command must be terminated with a <CR> character (ENTER key).
All
LOAD
functions require an
UPPER CASE
alphabetical character.
All
DISTANCE
or
MOTION CONTROL
functions require a
LOWER CASE
alphabetical
character.
NOTE: PRIOR TO DRIVING THE CS MACHINE USING ASCII COMMANDS 5
PARAMETERS MUST BE SET UP. THEY ARE SPEED, UPPER DISTANCE LIMIT,
LOWER DISTANCE LIMIT, UPPER FORCE LIMIT AND LOWER FORCE LIMIT.
17.4 ASCII COMMANDS TO DRIVE THE FORCE TESTER
The Force Tester can be configured to move the crosshead until a Load or Distance
measurement is reached.
The commands to achieve this are listed below: -
Send
Z
to zero the
LOAD
value.
Send
z
to zero the
DISTANCE
value
Send
H
???
to set the Upper
LOAD
limit (stop at this limit) for driving Upwards.
Send
G
-???
to set the Lower
LOAD
limit for driving Downwards.
Send
h
???
to set the Upper
DISTANCE
limit (stop at this limit) for driving Upwards.
Send
g
-???
to set the Lower
DISTANCE
limit for driving Downwards.
Send
e
???
to set the crosshead speed in in/min (inches per minute)
Send
u
to move the crosshead in the UP direction
Send
d
to move the crosshead in DOWN direction
Send
s
to stop the crosshead
The following shows how to set the machine to move up by 5 inches at 10 in/min.
Z (ENTER)
(Zero the Load)
z (ENTER) (Zero the Distance)
H225 (ENTER) (Set load upper limit to load cell capacity – Safety Limit)
G-225 (ENTER) (Set load lower limit to load cell capacity – Safety Limit)
h5 (ENTER) (Set the distance upper limit to 5 inches – Actual Limit)
g-1 (ENTER) (Set the distance lower limit to 1 inch – Safety Limit)
e10 (ENTER) (set the crosshead speed to 10in/min)
u (ENTER) (start the crosshead – this will stop at 5 inches)
Note that the s command to stop the crosshead is not required because the
crosshead will stop at 5 inches.
116 CS User Manual
Summary of Contents for Chatillon CS1100
Page 2: ...Part No 01 3953 MAY 2015 CS225 CS1100 Force Tester User Manual...
Page 50: ...Please refer to section 10 for details regarding test samples CS User Manual 49...
Page 126: ...CS User Manual 125...
Page 127: ...21 0 SPARE PARTS KITS SPK CS225 Sheet A 126 CS User Manual...
Page 128: ...SPK CS225 Sheet B CS User Manual 127...
Page 131: ...SPK CS1100 Sheet A 130 CS User Manual...