SECTION 3 – MAINTENANCE 3-5
5.
Update the calibration factor used in the display and controller. Follow the
manufacturers instructions for updating the factor. (CONFIG – IP menus).
Controller Calibration
The controller uses the same load cell signal that is used by the display. For this reason, they
use the same calibration factor.
1.
Obtain the new load cell calibration factor. Multiply the value in
mV/V
times the display
excitation voltage.
Output
Load
Voltage
Exec
Factor
Cal
=
*
2.
Enter the controller or display
Config
mode. Navigate to the
IP
menu and enter the
InP.H
value with the calculated load cell output corresponding to 50,000 Lbf (displayed
as 5000). This value must equal the calibration factor used by the load cell display.
3.
Exit the controller
Config
mode.
4.
Enter the controller
IP
menu. Adjust the
OFS.1
offset value until the displayed value
with zero load on the load cell is
0
. This may require some trial and error.
5.
Test the system to determine that the controller and display indicate the same load value.
Note that the value displayed by the controller must be multiplied by 10 to obtain Lbf.
The display indicates in units of Lbf directly.
Summary of Contents for Chandler Engineering 4207D
Page 8: ...P 4 PREFACE This page is intentionally left blank ...
Page 18: ...2 8 SECTION 2 OPERATING INSTRUCTIONS This page is intentionally left blank ...
Page 24: ...3 6 SECTION 3 MAINTENANCE This page is intentionally left blank ...
Page 28: ...5 2 SECTION 5 REPLACEMENT PARTS This page is intentionally left blank ...
Page 30: ...6 2 SECTION 6 DRAWINGS AND SCHEMATICS This page is intentionally left blank ...
Page 39: ...CHANDLER ENGINEERING ...
Page 40: ......