Maintenance
— 40
SPECTROLAB — 23.07.2020
9. Use a dry, grease-free cloth to
clean both sides of the spark
stand plate (1).
10. Use a dry, grease-free cloth to
clean the surface of the spark
stand body.
11. Use a dry, grease-free cloth or a
bottle brush with plastic bristles
to clean the light channel (2).
Fig. 19
12. Install the spark stand insert
The spark stand insert is
centered on the rear using
a contour (1).
Ensure that the spark stand
insert is flush with the top
edge (2) of the spark stand.
Fig. 20
13. Install the spark stand plate.
Fig. 21