Operation
— 30
SPECTROLAB — 23.07.2020
Standardizing the measuring device (optional)
Information
Prior to any standardization, the instrument must be iCALized. In addition, a
verification with certified standards must be performed after any standardization.
In order to standardize the instrument, proceed as follows:
1. Start the standardization function using the F7 function key.
2. Follow the instructions on the screen and perform the necessary
measurements with the profiling samples.
Use the "Finish" button or press the respective function key to finish the
respective sample.
After calculation of the standardization report, the standardization is
terminated automatically.
The instrument is now standardized.
Changing the base (multi-base instruments only)
When changing the base, the spark stand must be cleaned in order to avoid
memory effects, see page 37.
Change the spark stand insert, the electrode, and the electrode brush.
Preparation of the instrument for measuring another base is now complete.