2-14
3050-AP Moisture Analyzer
Problem
CPU hardware failure
Sample senor hardware fail-
ure
Action
Call AMETEK Service.
Replace sensor or call AME-
TEK Service.
This will occur during start-up
until the oven warms up. Call
AMETEK Service if problem
persists.
Check inlet and outlet pres-
sure. Call AMETEK Service if
problem persists.
Call AMETEK Service.
Check and/or replace dryer.
Call AMETEK Service if prob-
lem persists.
External temperature should
be 80
°
C or less. Call AMETEK
Service.
Replace dryer
Review alarm settings.
* Indicates System Alarm and Data Invalid Signal
Oven temperature is out of
tolerance.
Analyzer detected problem
with reference gas.
NOTE
Data valid contact opens on all alarms and stays closed during normal functions
and readings. An open data valid contact indicates verification is in process or
an alarm condition. Red LED flashes 11 times if an invalid condition occurs.
Sample flow rate too high or
too low
Battery needs to be replaced.
Excessive internal tempera-
ture.
Dryer failure Imminent
Moisture concentration is out
of user defined limits.
Table 2.1 LEDs and Alarms
Alarm Source /
LED Flashes per Cycle
Memory Failure*
1
Sample Sensor Failure*
2
Oven Temperature*
4
Flow Out of Tolerance*
5
Battery Low*
6
Reference Gas*
7
Enclosure Temperature
8
Dryer Alarm
10
Concentration Alarm
11
Summary of Contents for 3050-AP
Page 10: ...x 3050 AP Moisture Analyzer This page intentionally left blank...
Page 19: ...Instalation and Startup 2 3 Figure 2 1 Model 3050 AP front and side view...
Page 20: ...2 4 3050 AP Moisture Analyzer Figure 2 2 Model 3050 AP rear view...
Page 24: ...2 8 3050 AP Moisture Analyzer Figure 2 5 3050 AP Power and Signal Connections...
Page 25: ...Instalation and Startup 2 9 Figure 2 6 4 20 mA Output Wiring...
Page 26: ...2 10 3050 AP Moisture Analyzer Figure 2 7 RS 232 Wiring...
Page 27: ...Instalation and Startup 2 11 Figure 2 8 RS 485 Cables Multiple 3050 OLV s...
Page 28: ...2 12 3050 AP Moisture Analyzer Figure 2 9 RS 485 to RS 232 Conversion for host PC...
Page 46: ...5 4 3050 AP Moisture Analyzer This page intentionally left blank...