PN 903-8797, Rev A
Sub-Menu Item
Description
I1H
On the first analog output – V/I #1 – this is the full-scale (maximum)
WDP
temperature
(°C or °F) corresponding to the 20.00 mA output.
Note:
This parameter will be displayed only if an external water signal is provided and if
the
H2O DP:EXT
option is enabled from the
MonitorCFG
menu.
I2L
On the second analog output – V/I #2 – this is the low-scale (minimum)
HCDP
temperature
(°C or °F) corresponding to the 4.00 mA output.
I2H
On the second analog output – V/I #2 – this is the full-scale (maximum)
HCDP
temperature
(°C or °F) corresponding to the 20.00 mA output.
I3L
On the third analog output – V/I #3 – this is the low-scale (minimum)
WCT
– mg/m
3
(lb/MMcf),
PPMV, or PPMW – corresponding to the 4.00 mA output.
Note:
This parameter will be displayed only if an external water signal is provided and if
the
H2O CT:EXT
option is enabled.
I3H
On the third analog output – V/I #3 – this is the full-scale (maximum)
WCT
– mg/m
3
(lb/MMcf),
PPMV, or PPMW – corresponding to the 20.00 mA output.
Note:
This parameter will be displayed only if an external water signal is provided and if
the
H2O CT:EXT
option is enabled.
I4L
The fourth analog output – V/I #4 – is reserved for the low-scale (minimum) pressure
(KPA or PSI) corresponding to the 4.00 mA output.
This analog output is currently not used.
I4H
The fourth analog output – V/I #4 – is the full-scale (maximum) pressure (KPA or PSI)
corresponding to the 20.00 mA output.
This analog output is currently not used.
Chng PSCD1?
This prompt asks if you wish to change Passcode1. To change the password for the
ApplicCFG
menu, enter a new 4-character code, using any combination of the
, ‘
+
’, and ‘
–
’ buttons. After
entering the password, “
Chng PSCD1?
” will be displayed again. Enter the passcode a second
time.
Save AppCFG?
This prompt asks if you wish to permanently save the changes you made to the ApplicCFG
parameters. To save any changes you made, press the ‘
+
’ button. If the changes are not saved,
they will remain in effect until the analyzer is powered down or reset.
User Interface
|
3-17