PN 903-8797, Rev A
Working in the Application Configuration Menu
The
ApplicCFG
menu allows you to review and modify application- and
customer-specific parameters.
To view the
ApplicCFG
menu, press
until
ApplicCFG
appears. To view the
sub-menu items within this menu, press
. Enter the password at the “
Pass-
code1?
” prompt to view the
ApplicCFG
menu parameters.
The
ApplicCFG
sub-menu items (listed in their viewing order) and their de-
scriptions are:
Sub-Menu Item
Description
Passcode1?
This prompt asks you to enter Passcode1 to access the
ApplicCFG
menu (the factory default
password is ‘
- - - -
’). Enter this password to view and/or change the parameters in this menu.
WDPA
(Currently Not Used.)
This parameter displays the current Water Dewpoint temperature
OffSpec Alarm
setpoint (°C or °F). When the determined Water Dewpoint temperature is
greater than this setpoint temperature, a relay is activated and LED #7 on the User Interface is
turned On.
Note:
This signal will only be displayed if the
H2O DP:INT
option
is enabled from the
MonitorCFG
menu.
AMETEK does
not recommend enabling this except for diagnostic purposes only.
HCDPA
This is the current setpoint for the Hydrocarbon Dewpoint temperature Off-Spec Alarm (°C or
°F). When the determined Hydrocarbon Dewpoint temperature is greater than this setpoint
temperature, a relay is activated and LED #6 on the User Interface is turned On (if the Off-Spec
Alarm is enabled in the
Stat\History
menu). LED 1 is Off when the Off-Spec Alarm is enabled.
WCA
(Currently Not Used.)
This signal displays the current Water Content Off-Spec Alarm setpoint
(mg/m
3
, lb/MMcf, PPMV, PPMW). When the Water Content is greater than this setpoint
temperature, a relay is activated and LED #5 on the User Interface is turned On.
Note:
This signal will only be displayed if the
H2O DP:INT
option (plus water content) is
enabled from the
MonitorCFG
menu.
AMETEK does not recommend enabling
this except for diagnostic purposes only.
Purge
This parameter displays the current setting for the Measuring Cell Purge duration (minutes).
This setting is the length of time the solenoid valve is activated (open) to allow the sample gas
to flow through the Measuring Cell.
MTHi
This parameter displays the current
Mirror High
temperature setpoint (°C or °F) for the
Cooling Stage of the measuring cycle.
MTLo
This parameter displays the current
Mirror Low
temperature setpoint (°C or °F) for the Cooling
Stage of the measuring cycle.
FRate
This parameter displays the current Fast-per-minute cooling rate temperature (°C or °F) that
will be used during the Cooling Stage of the measuring cycle.
SRate
This parameter displays the current Slow-per-minute cooling rate temperature (°C or °F) that
will be used to cool the mirror during the Cooling Stage of the measuring cycle.
Above
This parameter displays the current setpoint for the Temperature Above the last
HCDP
temperature to switch the cooling rate (°C or °F). This value is also used to switch from the
Warming Stage to the Purging/Hold Stage.
A password is required
to access the ApplicCFG menu.
3-14 | 241CE II Hydrocarbon Dewpoint Analyzer