PN 903-8797, Rev A
Primary
Code
Error
Code
Alarm
Type
Condition / Description / Suggested Corrective Action
8
4
Warning
w Check XSCode
[Advice: Review-Extended-Status-Code]
XSCode alarm conditions may be present.
Corrective Action:
• Review and interpret current
XSCode
(
Stat\History
menu) for additional alarms. At-
tempt to correct any alarm conditions.
16
5
Warning
w Dirty Cell
[Possible-Cell-Contamination]
The Measuring Cell is contaminated with excessive condensates or other particulate.
Corrective Action:
• Review the
^HCMir
and
^WMir
history buffers (
Stat\History
menu) to see if either of
these parameters has changed recently or if one has a brightness level of over 400 at
the end of the Purging/Hold Stage. Review the End-of-Run type parameter
(
ApplicCFG : RunType
) and make sure it is set to HCDP.
• From the
RUN
menu check the
HCDP
temperature to see if it is particularly high. If so,
view the
ApplicCFG
menu and increase the
MTHi
setting and the
Purge
time, and
recheck the
HCDP
temperature.
• Ensure that the enclosure/shelter housing the analyzer is adequately heated. The
surrounding ambient temperature should be at least
15 °C (27 °F)
above the highest
expected HCDP temperature.
• If necessary clean the Measuring Cell and Filterblock Assemblies and replace all o-rings
and Membrane Filters. Also, inspect the sample lines and clean or replace them.
32
6
Fault
f H2O Optics
[Problem-In-Water-Side-Optical-Path]
T
here is a problem with the water-side optics within the Measuring Cell. An example
of this condition is if, at the beginning of the Purging/Hold Stage, the water-side
phototransistor signal is significantly higher or lower than normal.
Corrective Action:
•
Review and interpret the current
XSCode
(
Stat\History
menu) to determine the cur-
rent alarm conditions. Attempt to correct any alarms.
If necessary, contact AMETEK Service for assistance.
64
7
Fault
f HC Optics
[Problem-In-Hydrocarbon-Side-Optical-Path]
There is a problem with the hydrocarbon-side optics within the Measuring Cell.
An example of this condition is if, at the beginning of the Purging/Hold Stage, the
hydrocarbon-side phototransistor signal is significantly higher or lower than normal.
Corrective Action:
• Review and interpret the current
XSCode
(
Stat\History
menu) to determine the cur-
rent alarm conditions. Attempt to correct any alarm conditions.
If necessary, contact AMETEK Service for assistance.
4-48 | 241CE II Hydrocarbon Dewpoint Analyzer