12 Electrical and Signal Connections
All electrical work must be carried out by a competent person.
Power and signal cabling are the responsibility of the customer and should comply
with local regulations.
Cables should be selected of a conductor size and insulation suitable for the
purpose used.
The instrument and power leads should always be grounded/earthed.
Screened signal leads are recommended and a single ground termination made at
one end only.
Description
Requirement
Size
Digital
System OK
Alarms (mid and high)
Maintenance
Calibration Trigger
Standby input
8 core (1 or 2 cables
depending on instrument)
7/0.2 core
Diameter 7.5 to 10 mm
Current loop output
4 to 20mA
DC sourcing
Connect screen at
monitor
3 core screened (1 or
2 cables depending on
instrument)
7/0.2 core
Diameter 7.5 to 10 mm
*Blow-back control signal
to probe Stream A
1 core
7/0.2 core
Diameter 7.5 to 10 mm
*Not required if using antifreeze lines.