Installation
15
Application Guide
(For instructions on how to access alternate modes see 2.10.4)
2.10.1 Selecting the Calibration Mode for your application (Continued)
Application
Calibration Mode
Liquids and Slurries
Auto-Cal Mode #2
Granular /Solids with Bulk Density
greater than 20#’s per cubic foot
Manual Cal Mode #7
Granular/Solids with Bulk Density
Under 20#’s per cubic foot
Manual Cal Mode #8 (high
sensitivity)
Interface Measurement
Manual calibration Mode #7
Plugged Chute Indication for Solids
(Bulk density greater than 20 #’s per
cubic foot)
Manual calibration Mode #7
Plugged Chute Indication for Solids
(Bulk density under 20 #’s per cubic foot)
Manual calibration Mode #8
(high sensitivity)
2.10.2 Using ThePoint with Auto-Calibration mode #2
After ThePoint is installed in the vessel, simply apply power.
ThePoint electronic unit will auto calibrate.
Caution
The material being measured must be below the sensing element
when power is applied (Sensing element uncovered).
Note:
If power has been applied to ThePoint prior to installation (on a
test bench) or, if ThePoint is moved from one vessel to another,
RECAL
is necessary. RECAL allows the system's software to
capture the air capacitance generated by the sensing element
in tank.
Merely press and hold the RECAL button (shown in Figure
2-6) for five (5) seconds. After five seconds, ThePoint's two
LED’s flash for sixty seconds before the recalibration occurs.
(Removing power from the system while the LED's are flashing
will reset the memory immediately upon next power up).
Calibration is complete.
Summary of Contents for ThePoint Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 28: ...ThePoint Series 22 Section 3...
Page 36: ...Section 4...
Page 39: ...Control Drawings 33 Section 5 Control Drawings 5 1 FM FMc Control Drawings...
Page 40: ...ThePoint Series 34 5 1 FM FMc Control Drawings Continued...
Page 41: ...Control Drawings 35 5 1 FM FMc Control Drawings Continued...
Page 42: ...ThePoint Series 36 5 1 FM FMc Control Drawings Continued...
Page 43: ...Control Drawings 37 5 1 FM FMc Control Drawings Continued...
Page 44: ...ThePoint Series 38 5 1 FM FMc Control Drawings Continued...
Page 45: ...Control Drawings 39 5 1 FM FMc Control Drawings Continued...
Page 46: ...ThePoint Series 40 5 1 FM FMc Control Drawings Continued...
Page 47: ...Control Drawings 41 5 1 FM FMc Control Drawings Continued...
Page 48: ...ThePoint Series 42 5 1 FM FMc Control Drawings Continued...
Page 49: ...Control Drawings 43 5 1 FM FMc Control Drawings Continued...
Page 50: ...ThePoint Series 44 5 2 ATEX Control Drawings...
Page 51: ...Control Drawings 45 5 2 ATEX Control Drawings Continued...
Page 52: ...ThePoint Series 46 5 2 ATEX Control Drawings Continued...
Page 55: ...Control Drawings 49 5 2 ATEX Control Drawings Continued...
Page 56: ...ThePoint Series 50 5 3 Heavy Duty Spark Protection...
Page 57: ...Control Drawings 51 5 3 Heavy Duty Spark Protection Continued...
Page 58: ...ThePoint Series 52 5 4 Adding a Padded Capacitor...
Page 59: ...Control Drawings 53 5 4 Adding a Padded Capacitor Continued...
Page 60: ...ThePoint Series 54 5 4 Adding a Padded Capacitor Continued...
Page 61: ...Control Drawings 55 5 5 Dual Seal Assembly for 700 Series Sensing Elements...
Page 62: ...ThePoint Series 56 5 5 Dual Seal Assembly Continued...
Page 63: ...Control Drawings 57 5 5 Dual Seal Assembly Continued...
Page 64: ...Appendix A...
Page 69: ......