background image

OPERATING MANUAL FOR HEATSAN   

98-0177A 

  Page 6 

Tank Full Restart 

The tank full restart is the delay before the MRO unit starts when a tank full 
condition clears. This delay can be in minutes or in seconds. The TF Restart 
Setpoint selects seconds or minutes. 

Tank Full Override 

A timed tank full override can be initiated when the MRO unit is shut down due 
to a tank full condition. Pressing the Alarm Silence/Reset key for 3 seconds 
during a tank full condition will enable the tank full override. The RO will start 
and TF OVERRIDE 9 will show on the display. The number is the minutes 
remaining in the override timer. When the override times out, the unit will 
return to the tank full shut down condition. The TANK FULL OVERRIDE will 
divert all water to the drain, whether the water quality is good or bad coming 
into the RO. 

Pressure Fault 

If the pressure fault input becomes active (closes) and stays active for the 
delay programmed in the PF Delay Setpoint, the unit will shut down for a 
pressure fault. The display will show PRESS FAULT, the alarm lamp will flash 
and the audible alarm will sound. The pressure fault can be cleared by 
pressing the Alarm Silence/Reset key twice. 

Auto Reset 

If a pressure fault shut down occurs and the Auto Reset Setpoint is 
programmed to 0, the unit will remain shut down until manually reset. If the 
Auto Reset Setpoint is programmed to a value greater than 0, the unit will 
automatically clear the pressure fault and attempt to restart after this delay 
times out. 

Alarm Silence 

When a shut down occurs that causes the audible alarm to sound, the alarm 
can be silenced by pressing the Alarm Silence/Reset key once. The alarm will 
remain silenced for 3 minutes [180 seconds (AAMI RD62 standard)] when the 
Alarm Silence Setpoint is programmed to the factory default 3.  If the Alarm 
Silence Setpoint is programmed to a value other than 3, the alarm will 
resound after this delay times out. Pressing the Alarm Silence/Reset key will 
silence the alarm and reset this delay. 

Summary of Contents for 98-0177A

Page 1: ...OPERATION MANUAL FOR MRO CONTROLLER UPGRADE TO HEATSAN 98 0177A UPGRADE AN MRO FOR USE WITH HEATSAN 98 0177A 02 16 2015...

Page 2: ...OPERATING MANUAL FOR HEATSAN 98 0177A...

Page 3: ...OPERATION MANUAL FOR HEATSAN 98 0177A Page 3 CONTENTS PAGE 1 DOCUMENT HISTORY 1 2 CONTACT DETAILS 1 3 PREPARATION 2 4 CONTROLLER OPERATION 4 5 CONTROLLER STANDARD SETPOINTS 8...

Page 4: ...ription of Change A 17 02 15 Initial Release 2 CONTACT DETAILS For all service spares and consumables enquiries contact AmeriWater 1303 Stanley Ave Dayton Ohio 45404 Tel No 800 535 5585 Or your local...

Page 5: ...upgrade is complete Loosen the 2 retaining screws on the front of the control panel to gain access to the CPU Gently pry the CPU out of the socket being careful not to break any of the pins Discard t...

Page 6: ...o place Ensure that the chip is fully seated in the socket Return all power to the device Move the write protect jumper to the OFF position Power on the RO and access the program screen labelled as DI...

Page 7: ...mp starts OPERATING The RO unit is operating TANK FULL The unit is shut down due to a tank full condition TANK FULL 99 The unit is shut down due to a tank full condition If the number is blinking the...

Page 8: ...the switch is connected to the tank full high input When this switch has been active for 5 seconds the unit will shut down on tank full TANK FULL will show on the display When the tank full condition...

Page 9: ...s active for the delay programmed in the PF Delay Setpoint the unit will shut down for a pressure fault The display will show PRESS FAULT the alarm lamp will flash and the audible alarm will sound The...

Page 10: ...is warning will clear when the Conductivity drops below the Setpoint When the High Conductivity warning message is active the MRO will divert the PRODUCT WATER to drain through the Reject hose until t...

Page 11: ...ty in either S or PPM 0 S 1 PPM 0 1 0 Switch Select Selects if the switch inputs are normally open or normally closed 0 32 0 Temp UOM Selects display of the temperature in either F or C 0 F 1 C 0 1 0...

Page 12: ...inutes 3 Auto Reset When a pressure fault shut down is active the system will attempt to restart after this delay If set to 0 the system must be manually reset 0 99 minutes 0 Press Fault Delay The tim...

Reviews: