background image

 

 
 
 

 

 

 6 

 

AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o. 

 

5.

 

Using the conditioner 

Operation is reduced to few steps:  

»

 

Connect an ICP/IEPE standard sensor with the IN connector using a shielded cable, terminated 
with BNC plug 

»

 

Connect the conditioner output to measuring equipment (oscilloscope, data acquisition card, 
etc.) 

»

 

Connect an external power supply - necessary if the batteries haven’t been installed in the device 

»

 

Turn on the conditioner - switch should be placed to the ON position 

When the conditioner is turned on, the red LED labeled Power will lit, which means that the device is 
ready for use. 

6.

 

Troubleshooting 

The table contains the solution to some problems that the user may encounter during the operation of 
the conditioner. 

Problem 

Solution 

The unit will not turn on 

Check if the switch is set to the ON position. 

If the conditioner is battery powered, make sure that the batteries are installed in 
the device and that they are in working order. 

If an external power supply is used, make sure that the power adapter is connected 
correctly (with compatible polarity), in case of a regulated power supply make sure 
that it is set to supply 9V, check that the power supply is turned on. 

No signal on 
conditioner’s output 

Check if the ‘Sensor Error’ LED shines, if yes - check connection of the sensor to the 
conditioner and make sure that the sensor cable is OK - any breaks, cuts and 
abrasions qualify it to be replaced. Make sure that the sensor is operational. 

Output signal 
conditioner is noisy 
and/or disturbed 

When an external power supply is used, make sure that it is a transformer with 
linear stabilization. Using a power supply with pulse stabilization (inverter) may 
increase the noise level. 

Make sure the signal cable used to connect the sensor and the measuring 
equipment are shielded cables. 

Make sure that nearby there are no electrical devices turned on, that may interfere 
with the operation of the conditioner. 

Disconnect the AC adapter from the conditioner and power it up with batteries - in 
some cases, excessive noise can penetrate from the network through the AC 
adapter and affect the operation of the conditioner 

 

If the above table does not include the problem or given solution has been unsuccessful, contact AMC 
VIBRO service. 

 

Summary of Contents for AV CONDITIONER 1000

Page 1: ...ITIONER 1000 amc VIBRO Sp z o o Pilotow 2e 31 462 Krakow Poland T 48 12 362 97 60 S 48 12 362 97 63 info amcvibro pl KRS No 0000618618 REGON No 364497010 VAT No 6772403385 www amcvibro pl SINGLE CHANN...

Page 2: ...nts 1 Introduction 3 2 Technical data 3 3 Description of the AVC 1000 4 4 Power supply 4 4 1 Installing replacing batteries 5 4 2 Power supply 5 5 Using the conditioner 6 6 Troubleshooting 6 7 Recycle...

Page 3: ...frequency response very good linearity and low noise Is also equipped with an LED indicating open or short circuit in the sensor loop The device can be powered up either by an adapter or an internal b...

Page 4: ...ates the device 3 Red LED Sensor Error continuous light indicates open or short circuit 4 Red LED Power continuous light indicates operation 5 Red LED Low Batt continuous light indicates the need to r...

Page 5: ...ioner WARNING If the conditioner is not used for a long time remove the batteries to avoid leakage and damage 4 2 Power supply The unit can be powered with external power supply with 9V DC output volt...

Page 6: ...onnected correctly with compatible polarity in case of a regulated power supply make sure that it is set to supply 9V check that the power supply is turned on No signal on conditioner s output Check i...

Page 7: ...rcury cadmium hexavalent chromium polybrominated biphenyls polybrominated diphenyl ether or other battery related materials are limited to no more than trace amounts 7 2 Recycling Facilities When deco...