background image

 

 
 
 

 

 

 3 

 

AV CONDITIONER 1000 / ©AMC VIBRO Sp. z o.o. 

 

1.

 

Introduction 

AV CONDITIONER 1000

 (AVC 1000) is a universal conditioner designed to interface with sensors made 

in the ICP/IEPE standard. 

Thanks to its small size and weight, robust design, very good operational parameters and ease of use, it 
is ideally suited for carrying out measurements both in the laboratory and in the field. 

The  conditioner  is  equipped  with  input  circuits  enabling  powering  up  the  ICP/IEPE  sensor,  while 
maintaining wide frequency response, very good linearity and low noise. Is also equipped with an LED 
indicating open or short circuit in the sensor loop. 

The device can be powered up either by an adapter or an internal battery pack 4xAAA. 

2.

 

Technical data 

Parameters 

 

number of channels 

1

 

input type 

ICP/IEPE 24V DC / 2,4 mA

 

input / output connector 

BNC / BNC

 

input impedance 

100 kΩ 

output impedance 

120 Ω

 

gain 

1x (0 dB)

 

gain error 

< 0,1%

 

gain drift 

< 5 ppm /°C 

 

output range 

±10V

 

offset 

< 10 mV

 

SNR 

> 90 dB

 

frequency range (- 3 dB) 

0,16 Hz – 100 kHz *

 

power supply 

9V DC/100 mA  

 

batteries 

4 x AAA

 

dimensions 

60 x 100 x 30 mm

 

weight 

225 g

 

operating temperature 

0°C - 60°C

 

storage temperature 

-10°C – 70 °C (without batteries)

 

*) optional 500 kHz 

 

 

Summary of Contents for AV CONDITIONER 1000

Page 1: ...ITIONER 1000 amc VIBRO Sp z o o Pilotow 2e 31 462 Krakow Poland T 48 12 362 97 60 S 48 12 362 97 63 info amcvibro pl KRS No 0000618618 REGON No 364497010 VAT No 6772403385 www amcvibro pl SINGLE CHANN...

Page 2: ...nts 1 Introduction 3 2 Technical data 3 3 Description of the AVC 1000 4 4 Power supply 4 4 1 Installing replacing batteries 5 4 2 Power supply 5 5 Using the conditioner 6 6 Troubleshooting 6 7 Recycle...

Page 3: ...frequency response very good linearity and low noise Is also equipped with an LED indicating open or short circuit in the sensor loop The device can be powered up either by an adapter or an internal b...

Page 4: ...ates the device 3 Red LED Sensor Error continuous light indicates open or short circuit 4 Red LED Power continuous light indicates operation 5 Red LED Low Batt continuous light indicates the need to r...

Page 5: ...ioner WARNING If the conditioner is not used for a long time remove the batteries to avoid leakage and damage 4 2 Power supply The unit can be powered with external power supply with 9V DC output volt...

Page 6: ...onnected correctly with compatible polarity in case of a regulated power supply make sure that it is set to supply 9V check that the power supply is turned on No signal on conditioner s output Check i...

Page 7: ...rcury cadmium hexavalent chromium polybrominated biphenyls polybrominated diphenyl ether or other battery related materials are limited to no more than trace amounts 7 2 Recycling Facilities When deco...