10
EN
To use the mirror instead of the transmitted light lamp (I):
• Attach the mirror (AD) onto the mirror stand (AE)
• Unscrew the diffuser of the transmitted light lamp (I) from the base.
• Screw the mirror stand (AE) onto the base.
NOTICE
Using the mirror does not require electrical connection.
Operation
Switching on/off
• Connect the power cord (AF) to a power outlet and the other end to
the power socket (P).
NOTICE
To prevent the bulb in the lamp (I) from burning, turn the
brightness adjustment wheel (Q) setting to the lowest before turning on
or off.
• Flip the power switch (R) to “
I
” to turn on the lamp (I).
• After use, flip the power switch (R) to “
O
” to turn off.
Calculating magnification power
Total magnification power is calculated by multiplying the magnification
power of the eyepiece and the objective lens.
Eyepiece
magnification power
Objective lens
magnification power
Total magnification
power
10x (Z)
4x (V)
40x
10x (W)
100x
40x (X)
400x
100x (Y)
1000x
20x (AA)
4x (V)
80x
10x (W)
200x
40x (X)
800x
100x (Y)
2000x